LCD Multiparameter Patient Monitor ECG/ RESP/ SpO2/ 1IBP/ 2IBP/ 1T/ 2T/ NIBP/ REC/ INT. BATT.

Product not available

Add to Cart

GENERAL CHARACTERISTICS
FEATURES:
Menu, messages and instructions in English.
Single sweep speeds for circulatory and respiratory system: 12.5, 25 and 50 mm/s.
Visual and audible QRS indicator proceeding from ECG or Oxymeter.
When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
Three alarm levels: physiologicals, technicals and generals.
Upper and lower limits adjustable by the user.
Visual and audible alarm indicators, on screen and on top of the equipment.
Hotkey to activate the silent alarms for 2 minutes.
Alarms Turn Off.
User friendly menu by using the selector knob and hotkeys.
Waveform print-out on standard printer.
Analog outputs for ECG, RESP, IBP and a synchronization pulse.
Networking Connection for Centralized Monitoring.
WAIT Mode: stops monitoring, which allows changing to another patient without activating the alarms.
Screen menu: normal, 7 leads of ECG, Tabular trends or graphical trends, frozen and measuring of calipers, large numbers.

BASIC CHANNELS
ELECTROCARDIOGRAM
Heart rate: 20 to 300 1/min.
Gain selection: 1/4, 1/2, 1, 2 and 4 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
3 leads with 3 leads patient cable. Lead selection: DI, DII or DIII.
12 leads with 5 leads patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C1, C2, C3, C4, C5 or C6.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
Heart rate alarms, minimum and maximum: 20 to 300 1/min and off.
Lead off detection.
Pacemaker pulse detection.
Pacemaker pulse rejection in counting the heart rate:
Pulse rejected characteristic:
Amplitude: from ±2 mV to ±700 mV and pulse width: from 0.1 ms to 2.0 ms.

ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
Detects and classifies the following arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy, Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
Store up to 32 events and allows review and printing.
Measurement of the S-T segment: both the isoelectric point and the ST point are user adjustable (J-point +60 ms or +80 ms, or any other value in ms, from 1 to 120 ms).
Range: ±2.00 mV.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
- SpO2:
Range: 0% to 100%
Adult precision: 2% from 70% to 100%.
Neonate precision: 3% from 70% to 100%.
Motion precision: 3% from 70% to 100%.
Resolution: 1%.
Alarm: 0% to 100% and off.
Initial alarm: 85%.
- Pulse rate:
Range: 20 to 300 1/min.
Adult precision: ±3 1/min.
Neonate precision: ±3 1/min.
Motion precision: ±5 1/min.
Resolution: 1 1/min.
Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

RESPIRATORY WAVEFORM
Impedanciometric method.
Measurement selection: Transabdominal or Transtoraxic.
Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
Respiratory rate alarm, minimum and maximum: 5 to 170 1/min and off.

NON-INVASIVE BLOOD PRESSURE (NIBP)
Oscilometric method.
Use: ADULT or NEONATE.
Measurement mode: AUTOMATIC or MANUAL.
Measure unit: mmHg.
Resolution:
- Pressure: 1 mmHg.
- Pulse rate: 1 1/min.
Transductor accuracy: ±3 mmHg.
Ranges:
- PEDIATRIC/ADULT
Systolic pressure: 25 to 265 mmHg.
Diastolic pressure: 12 to 220 mmHg.
Mean pressure: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.
- NEONATE
Systolic pressure: 25 to 135 mmHg.
Diastolic pressure: 12 to 110 mmHg.
Mean pressure: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.
Alarm limits of: Systolic pressure from 20 to 255 mmHg and off.
Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
Protections:
Double protection from over-pressure.
Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
Protection for exceeding time of measurement.

INVASIVE BLOOD PRESSURE (IBP)
2 channels of invasive blood pressure.
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
Range: -30 to 300 mmHg.
Mode selection for Pressure Ranges: Automatic or Manual.
Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
Measure unit: mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

TEMPERATURE
2 channels of temperature.
Compatible with Thermistor Series YSI 400.
Measurement of body temperature or injectate temperature during measurement of Cardiac Output, from -5ºC to 45ºC.
Resolution: 0.1ºC. Accuracy: ±0.2ºC.
Temperature Alarm, minimum and maximum: -5ºC to 45ºC and off.

RECORDER
Hotkey to record command.
High resolution thermal array printer.
Recorder speed: 25 mm/s.
Recorded in real time up to 2 selectable waveforms between ECG, RESP, IBP, SpO2 and CO2 waveform.
Prints the table of cardiac output measurement and the waveform of blood temperature, during the injected.
Prints the header with date and time.
Print values of Sweep Speed, Lead, Gain, Respiratory Rate, Heart Rate, Systolic, Diastolic and Mean Pressure, SpO2 and Pulse Rate, EtCO2, FiCO2 and Respiratory Rate (CO2), according recorded waveform.

BATTERY
14.4V / 8.8Ah Li-ion battery.
Autonomy: 3.5 hours.
Battery state indication (empty, full, on charge) on screen.

ANY OTHER OPTIONAL CHANNEL SPECIFIED CAN BE ADDED IN THE FUTURE.

TECHNICHAL SPECIFICATIONS
15” LCD-TFT Color Monitor Resolution: 1024x768.
Greater angle of view.
Dimensions: Height: 345mm; Width: 335mm; Depth: 205mm.
Weight: 6Kg.
Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
Power Consumption: 60VA.

CERTIFICATIONS
Quality Standard (Certificates):
- ISO 13485.
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

Safety Standard (Certificates):
- UNE-EN60601-1.(Eq. IRAM 4220-1 and IEC 601-1).
- Electrical-shock protection: Class I – CF.
- UNE-EN60601-1-2 (Electromagnetic compatibility).
Authorized by A.N.M.A.T.: PM-1125-2.

ACCESSORIES
Included:
3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
SpO2 extension cable.
Cuff for NIBP.
Silicone hose for NIBP.
Pressure transducer adapter cable (DB9/G) to BD transducer.
Pressure transducer without adaptation B&D DTX plus TNF-R disposable.
Derivation box for parameters of Pressure 1, Pressure 2 and/or FiO2 with DB9M.
Temperature sensor Feas Electrónica - DB9M/G.
Derivation box for parameters of Temperature 1 and Temperature 2, DB9M connector.
Roll of Thermal paper of 50mm x 25m.
Power cord.
User manual.

The image is illustrative.


12 meses de garantía.