Monthly rent - Feas Electrónica's Automatic External Defibrillator (AED), model: Heart+ ResQ NT-381.C
GENERAL CHARACTERISTICS
Defibrillator
• Waveform: truncated exponential biphasic.
• 150 J (Adult, impedance 50 Ω).
• 50 J (Pediatric, impedance 50 Ω).
Charging time: 9 seconds (± 10 %).
• Shock method: manual button (flashes when ready).
• Mode conversion: Adult/Child toggle button.
(Child Mode delivers 50J Pediatric shock for patients weighing less than 25 Kg, approximately 12 months to 8 years old).

ECG
• ECG record: LEAD II.
• Analysis overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation.
• Analysis time: Approximately 7 seconds.

Battery Check
• Battery reserve level:
High: 3 bars.
Medium: 2 bars.
Low: 1 bar, red status led.
Critical: no bars, red status led.
• Battery status audible warnings.
Critical: Voice prompt states “Replace the battery”.

Data
• Stored data: ECG recordings, operation information (whether power is delivered or not), ambient noise, self diagnostic history.
• Memory capacity:
- ECG: Consecutively make records during analysis and CPR (once in every 5 s, up to 5000 s).
- Audio recording: 60 min in total (4 recording files of 15 min).
- Self-diagnosis: 3000 cases and results.
• Storage type: Internal memory.
• Access/search: “NT-ADV” software (for PC), via Bluethooth.

Audible and Visual Instructions
• After the equipment is switched on, the voice instructions will guide the operator through the resuscitation process, including the positioning of the patches, the indication of “Don’t touch the patient” (during the energy analysis and discharge process) and CPR (Cardio-Pulmonary Resuscitation) instructions.
• The CPR instructions tell the operator when to start Cardio-Pulmonary Resuscitation, according to the AHA Guide.

TECHNICAL SPECIFICATIONS
• Dimensions: 220 mm (W) x 282 mm (L) x 80 mm (H).
• Weight: 2 Kg (including battery and pads).
• Internal non-rechargeable Li-MnO2 battery - 12 Vdc/4200 mAh.
• Protection against ingress of liquids: IPX5 according to IEC60529.
• Temperature:
Storage: 15 ºC to 55 ºC.
Operation: 0 ºC to 45 ºC.
• Humidity:
Storage: 0 % to 95 % (without condensation).
Operation: 0 % to 90 % (without condensation).
• Altitude (Storage/Operation): 0 to 5000 meters.
• Drop test: 1.5 m according to IEC 60068-2-32.
• Quality standards:
ISO 13485 (certified).
Available A.N.M.A.T. 3266/2013: B.P.F.
Available A.N.M.A.T. 2319/02.
• Safety standards:
- IEC 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- IEC 60601-1-2 (Electromagnetic Compatibility).
- IEC 60601-1-6.
- IEC 60601-1-11.
- IEC 60601-2-4.

ACCESSORIES INCLUDED
- Electrode pads
• Shelf life: 2 (two) years.
• Single use.
• Cable length: 2.1 m (± 1 %).
• Applicable ages: All ages (12 months & older).
• Sealing: Airtight.
• Size: 125 mm x 95 mm.
• Pads placement:
Adult: anterior-lateral.
Pediatric: anterior-posterior (see positioning details on the package label).
- 1 Battery Pack
• Type: non-rechargeable Lithium-Manganese Dioxide (Li-MnO2).
• Voltage: 12 Vdc/4200 mAh.
• Standby mode life: Recommended five (5) years @ 25 ºC.
• Capacity: 180 shocks @ 25 ºC, 130 shocks @ 0 ºC.
• Measures: 22 mm (H) x 164 mm (W) x 54 mm (D).
• Weight: 260 g (±10 %).
- 1 User manual
- 1 bag
- Event Reading Software

OPTIONAL
USB/Bluetooth adapter.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Defibrillator Monitor 3850B Monophasic for simulation, Feas Electrónica
Simulation defibrillator. DOES NOT DISCHARGE ENERGY.

The Model 3850B Monophasic from Feas Electronica is an external defibrillator monitor with a monophasic waveform up to 360 J, selectable in steps.
It allows the monitoring and surveillance of ECG, with heart rate measurement, through the paddles, or cable to patient with 3 lead or 5 lead.

In addition, optionally, you can monitor pulse oximetry, graphing the plethysmographic curve on the screen and a pulse width indicator bar, indicating the SpO2 and pulse rate values. Both ECG and SpO2 monitoring allow the setting of alarms that allow you to monitor your heart rate, SpO2 value, and pulse rate.

It is of the portable type, powered by an internal rechargeable battery, a 90 Vac - 240 Vac, 50 Hz or 60 Hz mains supply or an external 12 Vdc source.
Optionally, it has a transcutaneous pacemaker that allows temporary electrical stimulation to the patient in a non-invasive way, in fixed mode (asynchronous) or on demand (synchronous), using electrodes.

Optionally, you can also have a thermal recorder for real-time recording of up to two curves, an ECG lead, and, optionally, the plethysmographic curve, manually when the operator decides or automatically during charging.


FEATURES
MONITOR
• Electrocardiogram monitor with LCD screen.
• Electrocardiogram signal acquired by cable to patient or by paddles.
• Derivations:
With 3-lead cable: DI, DII, DIII and PAL.
With 5-lead cable: DI, DII, DIII, aVR, aVF, aVL, Precordial and PAL.
• Cascade electrocardiogram.
• Digital indication of heart rate.
• Low battery indicator.
• Battery charging indicator light.
• Alarm silence key.
• Heart rate alarms from 20 [1 / min] to 250 [1 / min].
• ECG gain: 1/6, 1 / 4.1 / 2, 1 and 2 [cm / mV].

DEFIBRILLATOR
• Use of the Defibrillator in three steps:
1. Power on and power selection.
2. Loading.
3. Triggering of energy by means of push buttons on paddles or from keyboard.
• Step load selection key: 2, 3, 5, 7, 10, 20, 50,100, 200, 300 and 360 [Joules].
• Auditory and visual indication of end of charge.
• Indication on display of stored energy.
• Energy Trigger by simultaneously pressing both paddle pushbuttons or the “SHOCK”key.
• Internal charge cancellation:
It occurs when the "DISARM" key is pressed or when 60 seconds have elapsed from the end of the energy charge or when the energy selector switch is turned to the "MONITOR" or "OFF" position.
• It does not perform downloads but can, for simulation, be configured as direct or synchronized.
• Auditory and visual indication of synchronous mode.

REGISTER (OPTIONAL)
• Direct registration key.
• Automatic registration when the power charge process starts up to 20 seconds after pressing the Shock key.
• High resolution thermal recorder.
• Registration speed: 25 mm / s.
• Prints: Date and Time, Energy in Joules, EKC Curve, Gain, Heart Rate, Lead, Sweep Rate, Plethysmographic Curve, SpO2, Pulse Rate, Mode, Pacemaker Rate, Pacemaker Current.
• Paper: 50mm width, 45mm maximum diameter.
Accessory Included:
50 millimeter roll of paper.

TRANSCUTANEOUS PACEMAKER (OPTIONAL)
• External stimulation through electrodes.
• Frequency: 40 ppm to 180 ppm. (Selectable in steps of 5 ppm).
• Current: 10 mA to 200 mA. (Selectable in steps of 5 mA).
• Mode: Selectable, on demand or fixed.
• Pulse width: 20 ms ± 5%.
Accessories included:
Elastic belt for external Pacemaker electrode.
Pacemaker lead cable.
Pair of gray electrodes.

OXYGEN SATURATION (SpO2) (OPTIONAL)
SpO2 measurement:
• Range 0% to 100%.
• Accuracy:
Adult ± 2% between 70% and 100%.
Neonatal ± 3% between 70% and 100%.
In Motion ± 3% between 70% and 100%.
• Resolution: 1%.
• Percent oxygen saturation alarm:
Upper limit: 1 to 100.
Lower limit: 0 to 99.
• Measurement of pulse rate and alarms from 30 to 250 [1 / min] ± 3 [1 / min] and ± 5 [1 / min] in Motion.
• Resolution: 1 [1 / min].
Accessories included:
Oximetry Sensor Extender
Choice of adult or pediatric, clip-on or disposable oximetry sensor.

TECHNICAL SPECIFICATIONS
• Supply voltage: 90 Vac to 240 Vac - 50 Hz / 60 Hz, or 12 Vdc.
• Internal rechargeable NiMh battery - 12 V, 3500 mAh.
• Consumption: 10 W.
• Size: 295mm (width) x 200mm (height) x 345mm (depth).
• Weight: 5 Kg (Including external paddles, battery and recorder).

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monitoring central station model 5032C Feas Electrónica
The Monitoring central station model 5032C Feas Electronica is a centralized patient monitoring system, composed of a Central Processing Unit (CPU) with CENTRALC software, a color LCD monitor, a keyboard, a mouse, a printer and the necessary wiring for the communication with bedside devices. The Monitoring Central allows monitoring from 4 to 16 patients, through the connection with bedside monitors. Monitors supported include, but are not limited to, the following devices: LCD Multipair Patient Monitor, Defibrillator Monitor 3850B, and FEAS ELECTRONICA Vital Signs Monitors. The connection between the Bedside Monitor and the Monitoring Center allows communication which can be: Analog, Digital and Full Digital.

- Provides simultaneous monitoring of electrocardiogram with indication of heart rate, respiratory curve with indication of respiratory rate and invasive pressure curve, with indication of systolic, diastolic and mean pressure values of up to 16 different patients.
- Ideal for use in intensive care and coronary units.
- Observation of Waveforms: electrocardiogram, invasive blood pressure and respiratory.
- Digital communicaton with bedside feas ELECTRONICA monitors, for the reception of values of: ECG, Respiratory, SpO2, Pulse Rate, EtCO2, Invasive Blood Pressure, Temperature, Non-Invasive Blood Pressure, Fico, and Respiratory Rate.
- Selection of ECG and RESPIRATORY curve gains.
- Auto scale invasive pressure curve.
- Curve sweep speed selection.
- Audible and visual alarms by maximum and minimum, configurable for all the parameters; with volume control.
- Enter text on screen (Patient Case History).
- Freeze waveform of selected patient.
- Observe previous events.
- Trend graphs. Graphic or tabular trend of the last 72 hours, displayed on 1, 4, 8, 12, 24, 48 and 72 hours.
- High resolution printing of: saved waveforms and values for heart rate, systolic, diastolic and mean pressures, respiratory rate and trends for all the parameters.

TECHNICAL SPECIFICATIONS

Analogic Input Signals:
- ECG: 1V/mV.
- PRESSURES: 1V/100 mmHg.
- RESPIRATORY: 1V/3 Ohm.

Digital connection to bedside monitors:
- Display: Color SVGA Monitor Dimensions: 19" (Optional large size).
- CPU: Industrial PC Motherboard. Storage Solid State Disk.
- Output for connecting additional remote monitors.
- Printer: Laser black and white.
- AC Voltge input: 90V-/240V. ~ Frequency 50Hz/60Hz.

ACCESSORIES INCLUDED
- Keyboard, mouse and pad.
- Power cords for CPU, monitor and printer.
- User manual.

AREAS OF USE:
Intensive and intermediate care units, coronary care units, operating rooms, rehabilitation centers, others.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Electrophysiology and hemodynamic polygraph system mod. 1405 Feas Electrónica
The Polygraph is a system for conducting electrophysiology and hemodynamic studies.

The system is intended to be used in the acquisition, processing and recording of physiological signals. This includes surface electrocardiography (ECG) and intracavitary (IECG, for Electrophysiology only), heart rate, pulse oximetry (optional), invasive and non-invasive pressure (optional), cardiac output and hemodynamic calculations, now the last two for Hemodynamics only.

It is made up of a central data processing unit (CPU), a preamplifier, a color monitor, a laser printer, a keyboard, a mouse and a recording pedal. For Electrophysiology (if applicable), the Polygraph catheter extension box is added.

FEATURES

Hemodynamics
- 2 simultaneous ECG leads on screen, selectable from 12 leads using 10 electrode wires.
- Up to 4 Invasive Pressures.
- Automatic calculation of the fractional flow reserve (FFR) in real time, without the need for an independent (external) FFR analyzer. The FFR allows determining the hemodynamic severity of vascular lesions.
- Numerical and graphic display of the FFR value.
- Study storage.
- Cardiac output by thermodilution. Calculation of BSA, CO, CI, SV, SI.
- Selection of Catheter Constant K.
- Selection of the injection volume.
- Entry of patient data (height and weight).
- Measurement of the temperature of the blood and the injected.
- Curve of blood temperature variation.
- Messages on screen.
- Recording of studies, stores curves and calculations.
- Measurement of body or injection temperature (during cardiac output).
- Allows you to freeze curves and measure times and amplitudes using calipers.
- Printing of frozen curves or recorded studies.
• HEMODYNAMIC CALCULATIONS:
- Aortic valve area.
- Pulmonary valve area.
- Calculation of pulmonary and systemic vascular resistance.

ELECTROPHYSIOLOGY (OPTIONAL)
- 18 simultaneous endocavitational channels.
- Calipers on Screen: allow you to measure time between events of interest.
- 12 simultaneous surface ECG leads.
- Up to 4 Invasive Pressures.
- Tabular presentation of P.N.I.
- Split Screen: allows you to have a frozen reference image in one window, and in the other view the monitoring in real time.
- Quick access to the curve configuration window, where you can make a continuous selection of position, amplitude, filter selection and customize colors and labels of curves.
- Synchronized triggering of the curve sweep (trigger).
- Recording Studies on a hard disk, to be transported to other computers or personal computers using a pen drive.
- One SpO2 channel.
- Study printing with selection of color and line width.
- Catheters connection box and Cardiac Stimulator
- Review of recorded studies, with the possibility of modifying the gain, position, color and limits (of amplitude) of the curves.
- Quick access keys.

Pulse Oximetry (OPTIONAL)
- Plethysmographic curve, visualization of the oxygen saturation percentage.
- Low perfusion and movement signal measurement.
• SpO2:
- Range from 0% to 100%.
- Adult Accuracy: 2% from 70% to 100%.
- Neonatal Accuracy: 3% from 70% to 100%.
- Accuracy in motion: 3% from 70% to 100%.
- Resolution: 1%.
• Pulse Rate:
- Range from 20 [1 / min] to 300 [1 / min].
- Adult Accuracy: ± 3 [1 / min].
- Neonatal precision: ± 3 [1 / min].
- Accuracy in movement: ± 5 [1 / min].
- Resolution: 1 [1 / min].

Non-Invasive Pressure (OPTIONAL)
- Start of measurement manually or automatically at programmable intervals from 2 minutes to 90 minutes.
- Selection of Adult or Neonatal patient.
- Visualization of the last 1440 measurements.
- Printing of measurements in tabular form.
- Includes cuff.

TECHNICAL SPECIFICATIONS
• Screen: LCD monitor 19 ”
- Dimensions: 19 ”(larger size; optional)
- Resolution: 1024 x 768 dots
•CPU:
- Industrial PC motherboard.
- 160 Gb hard disk.
- Output to connect an additional remote monitor.
• Printing machine:
- Laser Technology.
- Resolution 1200 dpi.
• Power supply:
- 110V ~ / 220V ~.
- Frequency 50/60 Hz.

• Quality Standard:
- ISO9001 (certified).
- ISO13485 (certified).
- Disp. A.N.M.A.T. 3266/2013: B.P.F. (certificate).
- Disp. A.N.M.A.T. 2319/02 (certified).

• Safety Standard:
- UNE-EN60601-1 (Eq. IRAM 4220-1 and IEC 60601-1) Class I.
- Protection against electric shock: CF.
- Authorized by the A.N.M.A.T. PM 1125-23.

ACCESSORIES INCLUDED
- Preamplifier for Hemodynamic Studies.
- Interconnection cable from CPU to Preamplifier.
- Monitor, keyboard and mouse.
- Cable "Y" Minidin-M to 2 Minidin-F.
- Printing machine.
- Recording pedal with cable.
- CPU, monitor and printer power cables.
- Data cable to printer.
- Ground cable for interconnection between CPU and Preamplifier.
- Metal support for Preamplifier.
- Catheter extension box.
- Interconnection cable for catheter extension box.
- Mini equipment holder.
- TMM Bracket Mounting Clamp for TBC.
- Support plate for 4 transducers.
- Power strip with 5 shots.
- Disposable electrodes.
- 4 disposable pressure transducers.
- 4 interconnection cables from Preamplifier to disposable pressure transducers.
- Parameter derivation box for invasive pressure
- Injection temperature sensor.
- Patient cable with 3, 5 or 10 wires.
- Cardiac output catheter adapter cable.
- Oximeter sensor.
- Oximeter extension cable.
- Reus Adult Cuff.
- Silicone hose for PNI, 3m.
- Digital adapter for pressure guides (FFR).
- Volcano Pressure Guide Adapter Cable (FFR).
- Transport trolley (optional).
- User manual.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Patient Monitor, Multiparameter, LCD, Feas Electrónica
Feas Electrónica MULTIPAR LCD Patient Monitor is designed for monitoring the vital signs of patients in Intensive Therapies, Coronary Units, Operating Rooms, Emergency Rooms, and in a wide variety of clinical-hospital environments, where several vital signs need to be monitored of a patient simultaneously.

GENERAL CHARACTERISTICS
FUNCTIONS:
• Menu, messages and instructions in English.
• Single sweep speeds for circulatory and respiratory system: 12.5, 25 and 50 mm/s.
• Visual and audible QRS indicator proceeding from ECG or Oxymeter.
• When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
• Three alarm levels: physiologicals, technicals and generals.
• Upper and lower limits adjustable by the user.
• Visual and audible alarm indicators, on screen and on top of the equipment.
• Hotkey to activate the silent alarms for 2 minutes.
• Alarms Turn Off.
• User friendly menu by using the selector knob and hotkeys.
• Waveform print-out on standard printer.
• Analog outputs for ECG, RESP, IBP and a synchronization pulse.
• Networking Connection for Centralized Monitoring.
• WAIT Mode: stops monitoring, which allows changing to another patient without activating the alarms.
• Screen menu: normal, 7 leads of ECG, Tabular trends or graphical trends, frozen and measuring of calipers, large numbers.

BASIC CHANNELS
ELECTROCARDIOGRAM
• Heart rate: 20 to 300 1/min.
• Gain selection: 1/4, 1/2, 1, 2 and 4 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
• 3 leads with 3 wire patient cable. Lead selection: DI, DII or DIII.
12 leads with 5 wire patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C1, C2, C3, C4, C5 or C6.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
• Heart rate alarms, minimum and maximum: 20 to 300 1/min and off.
Lead off detection.
Pacemaker pulse detection.
• Pacemaker pulse rejection in counting the heart rate:
•Pulse rejected characteristic:
- Amplitude: from ±2 mV to ±700 mV
- pulse width: from 0.1 ms to 2.0 ms.

ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
• Detects and classifies the following arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy, Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
• Store up to 32 events and allows review and printing.
• Measurement of the S-T segment: both the isoelectric point and the ST point are user adjustable (J-point +60 ms or +80 ms, or any other value in ms, from 1 to 120 ms).
• Range: ±2.00 mV.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
• SpO2:
- Range: 35% to 100%
- Precision:
±2% from 80% to 100%.
±3% from 70% to 79%.
- Resolution: 1%.
- Alarm: 0% to 100% and off.
- Initial alarm: 85%.
• Pulse rate:
- Range: 25 to 250 1/min.
- Precision: ±2 1/min.
- Resolution: 1 1/min.
- Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

Accessories included:
Finger pulse oximeter sensor.

RESPIRATORY WAVEFORM
Impedanciometric method.
• Measurement selection: Transabdominal or Transtoraxic.
• Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
• Respiratory rate alarm, minimum and maximum: 5 to 170 1/min and off.

NON-INVASIVE BLOOD PRESSURE (NIBP)
Oscilometric method.
• Use: ADULT or NEONATE.
• Measurement mode: AUTOMATIC or MANUAL.
• Measure unit: mmHg.
• Resolution:
- Pressure: 1 mmHg.
- Pulse rate: 1 1/min.
• Transductor accuracy: ±3 mmHg.
• Ranges:
- PEDIATRIC/ADULT
Systolic pressure: 25 to 265 mmHg.
Diastolic pressure: 12 to 220 mmHg.
Mean pressure: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.
- NEONATE
Systolic pressure: 25 to 135 mmHg.
Diastolic pressure: 12 to 110 mmHg.
Mean pressure: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.
• Alarm limits of:
- Systolic pressure from 20 to 255 mmHg and off.
- Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
• Protections:Double protection from over-pressure.
- Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
- Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
- Protection for exceeding time of measurement.

Accessories included:
Cuff for Non Invasive Blood Pressure.
Hose for Non Invasive Blood Pressure.


INVASIVE BLOOD PRESSURE (IBP) (OPTIONAL)
1 or 2 channels of invasive blood pressure, according to customer's choice.
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
• Range: -30 to 300 mmHg.
• Mode selection for Pressure Ranges: Automatic or Manual.
• Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
• Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
• Measure unit: mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

TEMPERATURE (OPTIONAL)
2 channels of temperature.
Compatible with Thermistor Series YSI 400.
Measurement of body temperature or injectate temperature during measurement of Cardiac Output, from -5ºC to 45ºC.
• Resolution: 0.1ºC.
• Accuracy: ±0.2ºC.
• Temperature Alarm, minimum and maximum: -5ºC to 45ºC and off.

Accessories included:
Feas Electrónica temperature sensor - DB9M / G.
Derivation box for Temperature 1 and Temperature 2 parameters, DB9M connector (for two Temperature channels).

CARDIAC OUTPUT (OPTIONAL)
• Entry of patient parameters: Weight, height.
• Selection of:
- Injection volume: 3, 5 and 10 ml.
- Catheter Constant K or Computing Constant (CC)
- Measurement of the temperature in blood and injected.
• Calculation of:
- Body surface.
- CO: Volume in liters / min.
- SV: Stroke volume.
- CI: Cardiac index.
- SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement when injecting.

Accessories included:
Catheter adapter cable for cardiac output - DB9M/G
Derivation box for parameters Temperature 1 and Cardiac Output with DB9.
Derivative box for parameters Temperature 1, Temperature 2 and Cardiac Output with DB9.

SIDESTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 170 1 / min.
• Accuracy:
- CO2, ± 2 mmHg for CO2 <5%.
- <10% of reading for CO2> 5%.
• Resolution:
- CO2, 1 mmHg.
- Respiratory Rate, 1 1 / min.
• Alarm:
- FiCO2, EtCO2, minimum and maximum, 0 to 100 mmHg and off.
- Minimum and maximum Respiratory Rate, 0 to 170 1 / min and off.
Calibration: 2 zero points (ambient air) and gain (5% or 10%).
Compensation for temperature and atmospheric pressure, O2 and N2O.

Accessories included:
Type "T" nozzle for Feás Electrónica lateral flow capnography.
2.4 m patient line for Feás Electrónica lateral flow capnography
Moisture filter for lateral flow capnography, Feas Electrónica

SIDESTREAM CAPNOGRAPHY WITH ANESTHETIC GASES MEASUREMENT (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1/min.
• Accuracy: CO2, ±3%ABS.
• Resolution: CO2, 1 mmHg.
- Respiratory Rate, 1 1/min.
• Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
• Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
• Measurement of N2 O: from 0% to 100%.
•Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate:
- 70 to 120 ml/min for neonatal patients.
- 120 to 250 ml/min for adult patients.

Accessories included:
"T" type nozzle for sidestream Capnography, Feas Electrónica.
2,4 m patient line for sidestream Capnography, Feas Electrónica
Water trap for use in lateral flow capnography and anesthetic gases.

MAINSTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1 / min.
• Accuracy: ± 2 mmHg between 0 and 40 mmHg and ± 5% of reading 41 and 70 mmHg.
• Resolution: 1 mmHg.
• Calibration: only zero (ambient air).
Compensation for temperature and atmospheric pressure, O2 and N2O.
Warm-up time <15s, to start measuring, within 2 minutes and obtain measurement values according to specifications.
• Waveform response time: less than 60 ms.

Accessories included:
Mainstream Capnography sensor, Feas Electrónica
Nozzle for mainstream Capnography sensor

FiO2 (OPTIONAL)
Fraction of inspired Oxygen: from 0% to 100%.
Polarographic cell measurement principle.
Alarms minimum and maximum: 0% to 100% and off.

Accessories included:
Derivative box for pressure 1, pressure 2 and / or FiO2 parameters with DB9M.
FiO2 Sensor, Feas Electrónica.
Disposable Oxygen Sensor T-adapter for FiO2, Feas Electrónica
Spiral cable of Oxygen Sensor for FiO2, Feas Electrónica.

ANY OTHER OPTIONAL CHANNEL SPECIFIED CAN BE ADDED IN THE FUTURE.

TECHNICHAL SPECIFICATIONS
15” LCD-TFT Color Monitor
• Resolution: 1024x768.
Wide angle of visibility.
• Dimensions:
- Height: 345mm; Width: 335mm; Depth: 205mm.
• Weight: 6Kg.
• Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
• Power Consumption: 85VA.

CERTIFICATIONS
• Quality Standard (Certificates):
- ISO 13485.
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

• Safety Standard (Certificates):
- UNE-EN60601-1.(Eq. IRAM 4220-1 and IEC 601-1).
- Electrical-shock protection: Class I – CF.
- UNE-EN60601-1-2 (Electromagnetic compatibility).
Authorized by A.N.M.A.T.: PM 1125-2.

ACCESSORIES
Included:3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
SpO2 extension cable.
Cuff for NIBP.
Silicone hose for NIBP.
Temperature sensor Feas Electrónica - DB9M/G.
Derivation box for parameters of Temperature 1 and Temperature 2, DB9M connector.
FiO2 Sensor Feas Electrónica.
T Adapter for FiO2 sensor Feas Electrónica.
Spiral cable for FiO2 sensor Feas Electrónica.
Power cord.
User manual.
The image is illustrative.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Defibrillator Monitor 3850B Monophasic Feas Electrónica
The model 3850B Monophasic from FEAS ELECTRÓNICA is a monophasic waveform external defibrillator monitor up to 360J, selectable in steps, which allows defibrillation in asynchronous mode or synchronized with the R wave of the patient's ECG for the application of synchronized cardioversion.

It allows the monitoring and surveillance of ECG, with heart rate measurement, through the paddles, or cable to patient with 3 lead or 5 lead.

In addition, optionally, you can monitor pulse oximetry, graphing the plethysmographic curve on the screen and a pulse width indicator bar, indicating the SpO2 and pulse rate values. Both ECG and SpO2 monitoring allow the setting of alarms that allow monitoring of heart rate, SpO2 value and pulse rate.

It is of the portable type, powered by an internal rechargeable battery, a 90 Vac - 240 Vac, 50 Hz or 60 Hz mains supply or an external 12 Vdc source.
Optionally, it has a transcutaneous pacemaker that allows temporary electrical stimulation to the patient in a non-invasive way, in fixed mode (asynchronous) or on demand (synchronous), using electrodes.

Optionally, you can also have a thermal recorder for real-time recording of up to two curves, an ECG lead, and, optionally, the plethysmographic curve, manually when the operator decides or automatically during charging.

MONITOR
• Electrocardiogram monitor with LCD screen.
• Electrocardiogram signal acquired by cable to patient or by paddles.
• Derivations:
- With 3-lead cable: DI, DII, DIII and PAL.
- With 5-lead cable: DI, DII, DIII, aVR, aVF, aVL, precordial and PAL.
• Cascade electrocardiogram.
• Digital heart rate indicator.
• Low battery indicator.
• Battery charging indicator light.
• Alarm silence key.
• Heart rate alarms from 20 [beats / min] to 250 [beats / min].
• ECG gain: 1/6, 1/4, 1/2, 1 and 2 cm / mV.
• Equipment suitable for hospital use.

DEFIBRILLATOR
• Monophasic wave
• Use of the Defibrillator in three steps:
1- Power on and energy selection.
2- Loading.
3- Triggering of energy by means of pushbuttons on paddles or from keyboard.
• Charge selection key in steps: 2, 3, 5, 7, 10, 20, 50, 100, 200, 300 and 360 Joules.
• Auditory and visual indication of end of charge.
• Indication of stored energy on display.
• Triggering energy by simultaneously pressing both paddle pushbuttons or the “SCHOCK” key.
• The internal charge cancellation occurs when the "DISARM" key is pressed or when 60 seconds have elapsed from the end of the energy charge or when the energy selector switch is turned to the "MONITOR" or "OFF" position. Energy is discharged internally without being present in paddles.
• Direct or synchronized discharge.
• Auditory and visual indication of synchronous mode.
• Discharge capacity: 80 discharges at maximum energy, with the internal battery fully charged.
• Charging time at 360 J: less than 6 seconds.
• Connection to internal paddles or defibrillation pads.
• The equipment has a delivered energy test.

TRANSCUTANEOUS PACEMAKER (OPTIONAL)
• External stimulation through electrodes.
• Frequency: 40 ppm to 180 ppm (selectable in steps of 5 ppm).
• Current: 10 mA to 200 mA (selectable in steps of 5 mA).
• Mode: Selectable, on demand or fixed.
• Pulse width: 20 ms ± 5%.
Accessories included:
Elastic belt for external Pacemaker electrode.
Pacemaker lead cable.
Pair of gray electrodes.

OXYGEN SATURATION (SpO2) (OPTIONAL)
SpO2 measurement:
• Range: 0% to 100%.
• Accuracy:
Adult: ± 2% between 70% and 100%.
Neonatal: ± 3% between 70% and 100%.
In motion: ± 3% between 70% and 100%.
• Resolution: 1%.
• Percent oxygen saturation alarm
Upper limit: 1 to 100.
Lower limit: 0 to 99.
• Measurement of pulse rate and alarms from 30 to 250 [1 / min] ± 3 [1 / min] and ± 5 [1 / min] in motion.
• Resolution: 1 [1 / min].
Accessories included:
Oximetry Sensor Extender
Choice of adult or pediatric, clip-on or disposable oximetry sensor.

REGISTER (OPTIONAL)
• Direct registration key.
• Automatic registration when the power charge process starts up to 20 seconds after pressing the Shock key.
• High resolution thermal recorder.
• Registration speed: 25 mm / s.
• Prints: date and time, energy in Joules, EKG curve, Gain, Heart rate, Lead, Sweep speed, Plethysmographic curve, SpO2, Pulse rate, Mode, Pacemaker rate, Pacemaker current.
• Paper: width of 50 mm and maximum diameter of 45 mm.
Accessory Included:
50 millimeter roll of paper.

TECHNICAL SPECIFICATIONS
• Supply voltage: 90Vac to 240Vac - 50/60 Hz or 12 Vdc.
• Internal rechargeable NiMh battery - 12V, 3500 mAh.
• Consumption 180 W (7W during standby).
• Size: 295mm (width) x 200mm (height) x 345mm (length).
• Weight: 6 kg (including external paddles, battery and recorder).

CERTIFICATIONS
• Quality standards:
- ISO13485 (certified).
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

• Safety rules:
- EN 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- EN 60601-1-2 (Electromagnetic Compatibility).
- EN 60601-1-6.
- EN 60601-1-8.
- EN 60601-2-4.
- EN 60601-2-27.
- EN 60601-2-49.
- ISO 80601-2-61.
Authorized by the A.N.M.A.T. PM-1125-4.

ACCESSORIES INCLUDED
External paddle set for adults.
Power cable for external +12 Vdc input.
Line power cable.
Patient cable with 3 or 5 wires for ECG.
10 disposable electrodes.
User manual.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Defibrillator Monitor 3850B Biphasic Feas Electrónica
Feas Electronica Model 3850B Biphasic is an external defibrillator monitor with a rectilinear biphasic waveform of up to 200J, selectable in steps, which allows defibrillation in asynchronous or synchronized mode with the R wave of the patient's ECG for the application of synchronized cardioversion.
It allows the monitoring and surveillance of ECG, with heart rate measurement, through the paddles, or cable to patient with 3 lead or 5 lead.

In addition, optionally, you can monitor pulse oximetry, graphing the plethysmographic curve on the screen and a pulse width indicator bar, indicating the SpO2 and pulse rate values. Both ECG and SpO2 monitoring allow the setting of alarms that allow monitoring of heart rate, SpO2 value and pulse rate.

It is of the portable type, powered by an internal rechargeable battery, a 90 Vac - 240 Vac, 50 Hz or 60 Hz mains supply or an external 12 Vdc source.
Optionally, it has a transcutaneous pacemaker that allows temporary electrical stimulation to the patient in a non-invasive way, in fixed mode (asynchronous) or on demand (synchronous), using electrodes.

Optionally, it can also have a thermal recorder for real-time recording of up to two curves, an ECG lead, and optionally the plethysmographic curve, manually when the operator decides or automatically during charging.

FEATURES

MONITOR
• Electrocardiogram monitor with LCD screen.
• Electrocardiogram signal acquired through 3 or 5 leads patient cable or through paddles.
• Leads selection:
- With 3 lead cable: DI, DII, DIII and PAL.
- With 5 lead cable: DI, DII, DIII, aVR, aVF, aVL, precordial and PAL.
• ECG Gain: 1/6, 1/4, 1/2, 1 and 2 cm / mV.
• ECG waveform cascade.
• Digital indication of heart rate.
• Low battery indicator.
• Luminous Indicator of battery charging.
• Alarms silence key.
• Alarms for Heart Rate from 20 ppm to 250ppm.

DEFIBRILLATOR
• Biphasic waveform.
• Defibrillator use in three steps:
1- Turn on and select energy.
2- Load.
3- Trigger energy discharge by means of push buttons on paddles or from the keyboard.
• Selector switch for charge selection by steps: 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200 Joules.
• Visual and audible indication of charging process completed.
• Stored energy indication on display.
• Energy discharge through paddles by pushing, simultaneously, both pushbuttons on paddles.
• Charge is cancelled internally: When the key DISARM is pressed or when 60 seconds have elapsed since charge process was completed or when selector switch is turned to MONITOR or OFF.
In these cases the energy is discharged internally without reaching the paddles.
• Direct or synchronized discharge.
• Visual and audible indication of synchronized mode.
• Discharge capacity: 150 discharges at maximum energy, with fully charged internal battery.
• Charging time to 200 J, less than 6 seconds.
• The equipment has a power test delivered.

RECORDER (OPTIONAL)
• Direct access key for record command.
• Automatic registration when the power charge process starts up to 20 seconds after pressing the Shock key.
• High resolution thermal array printer.
• Registration speed: 25 mm / s.
• Prints: date and time, energy in joules, Elctrocardiogram waveform, Gain, Heart Rate, Lead, Sweep Speed, Plethysmographics waveform, SPO2, Pulse Rate, Mode, Pacemaker Rate, Pacemaker Current.
• Paper: Width: 50 mm and Diameter Max: 45 mm.
Accessory Included:
50 millimeter roll of paper.

OXYGEN SATURATION (SpO2) (OPTIONAL)
SPO2 measurement:
• Range: 0 to 100%.
• Precision:
Adult ± 2% from 70% and 100%.
Neonatal: ± 3% from 70% and 100%.
In movement: ± 3% from 70% and 100%.
• Resolution: 1%.
• Percentual oxygen saturation alarm:
Upper limit: 1 to100.
Lower limit: 0 to 99.
• Pulse Rate and alarms measurement from 30[1/min] to 250 [1/min] ± 3[1/min] and ± 5 [1/min] with movement.
• Resolution: 1 [1/min].
Accessories included:
Oximetry Sensor Extender
Choice of adult or pediatric, clip-on or disposable oximetry sensor.

TRANSCUTANEOUS PACEMAKER (OPTIONAL)
• External stimulation through electrodes.
• Frequency: 40 ppm to 180 ppm (selectable in steps of 5 ppm).
• Current: 10 mA to 200 mA (selectable in steps of 5 mA).
• Mode: Selectable, on demand or fixed.
• Pulse width: 20 ms ± 5%.
Accessories included:
Elastic belt for external Pacemaker electrode.
Pacemaker lead cable.
Pair of gray electrodes.

TECHNICAL SPECIFICATIONS
• Power input: 90Vac ~ to 240Vac - 50Hz/60Hz or 12 Vcc.
• Rechargeable internal battery, NiMH - 12V, 3500 mAh.
• Power requirement: 180 W (7W during standby).
• Size: 295mm (width) x 200mm (height) x 345mm (length).
• Weight: 6 kg (including external paddles, battery and recorder).

CERTIFICATIONS
• Quality standards:
- CE Marking, Dir. 93/42 / EEC (certificate).
- ISO13485 (certificate).
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.
• Safety rules:
- EN 60601-1 (Eq IRAM 4220-1 and IEC 601-1).
- EN 60601-1-2 (EMC).
- EN 60601-1-6.
- EN 60601-1-8.
- EN 60601-2-4.
- EN 60601-2-27.
- EN 60601-2-49.
- ISO 80601-2-61.
Authorized by the A.N.M.A.T. PM-1125-15.

INCLUDED ACCESSORIES
External paddles for adults and pediatric with coil cable.
Power cable for external +12 power supply.
AC Power cord.
3 or 5 wire cable for patient for ECG.
Disposable electrodes.
User's manual.

The image is illustrative.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Keyboard for Comen electrocardiographs, models CM100 and CM300
Keyboard for Comen electrocardiographs, models CM100 and CM300


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Pediatric pulse oximeter with Clamp and DB9M/G, Feas Electrónica 3178
Pediatric pulse oximeter with Clamp and DB9M/G, Feas Electrónica 3178

The reusable finger sensor is used in oximetry monitoring systems to non-invasively determine Oxygen saturation (SpO2), pulse rate, and pulse plethysmographic waves.

The reusable pediatric finger sensor is recommended for:
• Patients weighing more than 5Kg and less than 45Kg.
• Patients with limited activity.
• For sampling.

The reusable pediatric finger sensor is not recommended for long-term use. For long-term patient monitoring, use disposable single-patient or wrap-around type sensors.
It has a length of 1 meter.
Protection against liquids IPX1.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Adult Oximeter Finger Sensor DB9M/G Feas Electrónica
Adult Oximeter Finger Sensor DB9M/G Feas Electrónica

The Oximeter sensor is applicable to adult patients (weighing more than 30kg), for non-invasive measurement by continuous sampling of arterial oxygen saturation and pulse rate, in Oximetry Monitoring Systems. This sensor is designed to be used exclusively in Feas Electronica equipment.

It has a length of 2.5 meters.
Protection against liquids: IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Adult pulse oximeter sensor with clamp and DB9M/G 3444, Feas Electrónica
Adult pulse oximeter sensor with clamp and DB9M/G 3444, Feas Electrónica

The reusable finger sensor Feas ELECTRONICA ref. 5043 is used to non-invasively determine oxygen saturation (SpO2), pulse rate, and pulse plethysmographic waves. It is indicated for:
• Patients who weigh more than 15Kg.
• Patients with limited activity.

The reusable finger sensor is not recommended for prolonged use, unless the integrity of the skin is checked frequently and the application site is changed accordingly. For long-term patient monitoring, use either single-patient disposable sensors or Velcro-strap sensors.
It has a length of 1 meter.
Protection against liquids IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Neonatal oximeter sensor DB9M/G Feas Electrónica 3026
Neonatal oximeter sensor DB9M/G Feas Electronica 3026.

This sensor is used in oximetry monitoring systems to non-invasively determine Percent Oxygen Saturation (SpO2), pulse rate, and pulse plethysmographic waves.
Its use is indicated for patients of less than 3kg.

It has a length of 40 cm.
Protection against liquids IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Oxygen sensor with tongue clips for veterinary use. D9MG, Feas Electrónica. 551A
Oxygen sensor with tongue clips for veterinary use. D9MG, Feas Electronica.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Paper compartment door for Comen ECG, model CM300
Paper compartment door for Comen ECG, model CM300

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Thermoformed, mechanized and milled plastic paddle holder for Defibrillator 3850B
Thermoformed, mechanized and milled plastic paddle holder for 3850B Defibrillator.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Polycarbonate NIBP parameter, revision 01, front of Multipar LCD version 7.x
Self-adhesive polycarbonate for multipair monitor of the NIBP parameter connector.
Self-adhesive polycarbonate of "SILENCE" key for Multipar keyboard
Self-adhesive polycarbonate of "SILENCE" key for Multipar keyboard

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Self-adhesive polycarbonate of key "ESC" for keyboard of Multipar
Self-adhesive polycarbonate of key "ESC" for keyboard of Multipar

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Gray nylon KNOB FOR .250" electronic switch PUSHBUTTON
Grayhill trim knob. Replacement for rotary knob of the Multipair Monitor

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.