Patient Monitor Model PM9000 Feas Electrónica
MULTIPARAMETER PATIENT MONITOR Model: PM9000
GENERAL CHARACTERISTICS
Features:
Menu, messages and instructions in English or Spanish.
Single sweep speeds for circulatory (25mm/s y 50mm/s) and respiratory system (6,25 mm/s, 12,5 mm/s, 25 mm/s y 50 mm/s).
Visual and audible QRS indicator proceeding from ECG or Oxymeter.
When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
Physiological and technical alarms.
Three alarm priority levels: low, medium and high priority.
Upper and lower limits adjustable by the user.
Visual and audible alarm indicators, on screen and on top of the equipment.
Hotkey to pause visual or audible alarms for 2 minutes.
Visual or audible alarms Turn Off function.
User friendly menu by using the trim knob and hotkeys.
Networking Connection for Centralized Monitoring.
Screen menu: - Normal.
- 7 leads of ECG.
- Tabular trends or graphical trends.
- Big numbers
- OxyCRG
BASIC CHANNELS
Electrocardiogram (ECG)
Heart rate: Adult 15 to 300 1/min. Pediatric 15 a 350 1/min
Gain selection: 0.25, 0.5, 1 and 2 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
3 leads with 3 leads patient cable. Lead selection: DI, DII or DIII.
7 leads with 5 leads patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
Heart rate alarms: Lower and upper (medium priority).
Lower and upper extreme (high priority) and off.
Lead off detection.
Pacemaker pulse indication.
Pacemaker pulse rejection in counting the heart rate:
Pulse rejected characteristic: -Amplitude: from ±2 mV to ±700 mV.
-Pulse width: from 0.1 ms to 2.0 ms.
Analysis and classification of arrhythmias
Detected and classified the follow arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy,
Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
Measurement of the S-T segment: both the isoelectric point and the ST point areuser adjustable.
Range: ±2.00 mV. PVCs per minute measurement.

Respiratory waveform (RESP)
Impedancemetry Method.
Measurement selection: Transabdominal (II-F) o Transthoracic (I-L).
Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
Respiratory rate alarm, minimum and maximum: 5 to 170 1/min.

Oxygen saturation (SpO2)
Plethysmographic waveform, Display of percentual oxygen saturation.
Measurement signal low perfusion and motion.
SpO2:
Range: 0% to 100%.
Adult precision: 2% from 70% to 100%.
Neonate precision: 3% from 70% to 100%.
Motion precision: 3% from 70% to 100%.
Resolution: 1%.
Alarm: 0% to 100% and off.
Initial alarm: 85%.
Pulse rate:
Range: 20 to 250 1/min.
Adult precision: ±3 1/min.
Neonate precision: ±3 1/min.
Motion precision: ±5 1/min.
Resolution: 1 1/min.
Alarm: 20 to 250 1/min and off.

Two channels of invasive blood pressure (IBP)
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
Range: -30 to 300 mmHg.
Mode selection for Pressure Ranges: Automatic or Manual.
Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
Technical alarm due to transducer failure and disconnected catheter.
Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
Units in mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

Two temperature channels
Compatible with Thermistor Series YSI 400.
Measurement of body temperature (25°C a 50°C) or injectate temperature during measurement of Cardiac Output, from (-10ºC to 25ºC).
Resolution: 0.1ºC. Accuracy: ±0.2ºC.
Temperature Alarm, minimum and maximum:(25ºC to 50ºC) and off.

Cardiac output (CO)
Input of patient parameters: Weight, height.
Selection of: Injectate volume: 3, 5 and 10 ml.
Catheter Constant K or Computing Constant (CC).
Measurement of temperature in blood and injected.
Calculation of: Body surface.
CO: Volume in liters/min.
SV: Systolic volume.
CI: Cardiac index.
SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement, at perform the injection.
Stores 8 measurements.
Blood temperature: Range 15°C to 45°C, resolution 0.01°C and accuracy ±0.3°C.
Injected temperature: Range -10°C to 25°C, resolution 0.1°C and accuracy ±0.4°C.

Non-invasive blood pressure (NIBP)
Oscillometric method
Use: ADULT/PEDIATRIC or NEONATE.
Measurement mode: AUTOMATIC or MANUAL.
Unit in mmHg.
Resolution:
Pressure: 1 mmHg.
Pulse rate: 1 1/min.
Transductor accuracy: ±3 mmHg or ±2%, whichever is greater.
Ranges:
ADULT/PEDIATRIC
Pressure: Systolic: 25 to 265 mmHg.
Diastolic: 12 to 220 mmHg.
Mean: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.

NEONATE
Pressure: Systolic: 25 to 135 mmHg.
Diastolic: 12 to 110 mmHg.
Mean: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.

Alarm limits of: Systolic pressure from 20 to 255 mmHg and off.
Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
Protections: Double protection from over-pressure.
Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
Protection for exceeding the time of measurement.

OPTIONALS
FiO2
Fraction of inspired Oxygen: from 0% to 100%.
Galvanic cell measurement principle.
Alarms minimum and maximum: 18% to 100% and off.

Recorder
Direct access key for record command.
High resolution thermal array printer.
Recorder speed: 25 mm/s.
Recorded in real time up to 2 selectable waveforms between ECG, RESP, IBP,
SpO2 and CO2 waveform.
Prints the header with date and time.
Print values of Sweep Speed, Lead, Gain, Respiratory Rate, Heart Rate,
Systolic, Diastolic and Mean Pressure, SpO2 and Pulse Rate, EtCO2, FiCO2
and Respiratory Rate (CO2), according recorded waveform.

Capnography

Option 1: Sidestream
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate: 4 to 150 1/min.
Accuracy: from 0 mmHg to 38 mmHg: ±2.0 mmHg
from 39 mmHg to 70 mmHg: ±5 mmHg
from 71 mmHg to 100 mmHg: ±8 mmHg
Resolution: CO2, 1mmHg.
Respiratory Rate, 1 1/min.
Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100mmHg and off.
Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
Calibration: Two zero points -room air and gains (5% or 10%)-.

Option 2: Sidestream with anesthetic gases measurement
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate, 4 to 150 1/min.
Accuracy: CO2, ±3%ABS.
Resolution: CO2, 1 mmHg.
Respiratory Rate, 1 1/min.
Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
Measurement of expired and inspired O2: from 0% to 100%, Resolution: 1%.
Measurement of N2O: from 0% to 100%.
Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,
Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate: 70 to 120 ml/min for neonatal patients.
120 to 250 ml/min for adult patients.

Option 3: Mainstream
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate, 4 to 170 1/min.
Accuracy: ±2 mmHg from 0 to 40 mmHg.
±5% of reading from 41 to 70 mmHg.
Resolution: 1 mmHg.
Calibration: only zero (room air).
Compensation for temperature and atmospheric pressure, O2 y N2O.
Warm-up time <15s, to start measuring, within 2 minutes to obtain values consistent with the specifications.
Waveform response time: less than 60ms.

Touchscreen

ACCESSORIES
Included:
3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
Pressure Transducer Adaptor (2 units).
Pressure transducer (2 units).
Temperature sensor.
Power cord.
User manual.

TECHNICAL SPECIFICATIONS
15” LCD Color Monitor. Resolution: 1024 x 768. Greater angle of view.
Dimensions: Height: 345mm; Width: 335mm; Depth: 205mm.Weight: 6 Kg.
Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
Power Consumption: 65VA.
Internal Battery: 14.4V / 8.8Ah Li-ion battery. Autonomy: 3.5 hours.
Battery state indication (empty, full, on charge) on screen.
Quality Standard: ISO 13485, Disp. A.N.M.A.T. 3266/2013: B.P.F. and
Disp. A.N.M.A.T. 2319/02, certified.
Safety Standard: UNE-EN60601-1. (Eq. IRAM 4220-1 and IEC 601-1) Electrical-
shock protection: Class I - CF, UNE-EN60601-1-2 (EMC), certified.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - LCD Multiparameter Patient Monitor ECG/RESP/SPO2/PNI
Includes equipment rental. For different rental periods, please consult

Feas Electrónica MULTIPAR LCD Patient Monitor is designed for monitoring the vital signs of patients in Intensive Therapies, Coronary Units, Operating Rooms, Emergency Rooms, and in a wide variety of clinical-hospital environments, where several vital signs need to be monitored of a patient simultaneously.

GENERAL CHARACTERISTICS
FUNCTIONS:
• Menu, messages and instructions in English.
• Single sweep speeds for circulatory and respiratory system: 12.5, 25 and 50 mm/s.
• Visual and audible QRS indicator proceeding from ECG or Oxymeter.
• When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
• Three alarm levels: physiologicals, technicals and generals.
• Upper and lower limits adjustable by the user.
• Visual and audible alarm indicators, on screen and on top of the equipment.
• Hotkey to activate the silent alarms for 2 minutes.
• Alarms Turn Off.
• User friendly menu by using the selector knob and hotkeys.
• Waveform print-out on standard printer.
• Analog outputs for ECG, RESP, IBP and a synchronization pulse.
• Networking Connection for Centralized Monitoring.
• WAIT Mode: stops monitoring, which allows changing to another patient without activating the alarms.
• Screen menu: normal, 7 leads of ECG, Tabular trends or graphical trends, frozen and measuring of calipers, large numbers.

BASIC CHANNELS
ELECTROCARDIOGRAM
• Heart rate: 20 to 300 1/min.
• Gain selection: 1/4, 1/2, 1, 2 and 4 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
• 3 leads with 3 wire patient cable. Lead selection: DI, DII or DIII.
12 leads with 5 wire patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C1, C2, C3, C4, C5 or C6.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
• Heart rate alarms, minimum and maximum: 20 to 300 1/min and off.
Lead off detection.
Pacemaker pulse detection.
• Pacemaker pulse rejection in counting the heart rate:
•Pulse rejected characteristic:
- Amplitude: from ±2 mV to ±700 mV
- pulse width: from 0.1 ms to 2.0 ms.

ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
• Detects and classifies the following arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy, Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
• Store up to 32 events and allows review and printing.
• Measurement of the S-T segment: both the isoelectric point and the ST point are user adjustable (J-point +60 ms or +80 ms, or any other value in ms, from 1 to 120 ms).
• Range: ±2.00 mV.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
• SpO2:
- Range: 35% to 100%
- Precision:
±2% from 80% to 100%.
±3% from 70% to 79%.
- Resolution: 1%.
- Alarm: 0% to 100% and off.
- Initial alarm: 85%.
• Pulse rate:
- Range: 25 to 250 1/min.
- Precision: ±2 1/min.
- Resolution: 1 1/min.
- Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

Accessories included:
Finger pulse oximeter sensor.

RESPIRATORY WAVEFORM
Impedanciometric method.
• Measurement selection: Transabdominal or Transtoraxic.
• Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
• Respiratory rate alarm, minimum and maximum: 5 to 170 1/min and off.

NON-INVASIVE BLOOD PRESSURE (NIBP)
Oscilometric method.
• Use: ADULT or NEONATE.
• Measurement mode: AUTOMATIC or MANUAL.
• Measure unit: mmHg.
• Resolution:
- Pressure: 1 mmHg.
- Pulse rate: 1 1/min.
• Transductor accuracy: ±3 mmHg.
• Ranges:
- PEDIATRIC/ADULT
Systolic pressure: 25 to 265 mmHg.
Diastolic pressure: 12 to 220 mmHg.
Mean pressure: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.
- NEONATE
Systolic pressure: 25 to 135 mmHg.
Diastolic pressure: 12 to 110 mmHg.
Mean pressure: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.
• Alarm limits of:
- Systolic pressure from 20 to 255 mmHg and off.
- Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
• Protections:Double protection from over-pressure.
- Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
- Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
- Protection for exceeding time of measurement.

Accessories included:
Cuff for Non Invasive Blood Pressure.
Hose for Non Invasive Blood Pressure.


INVASIVE BLOOD PRESSURE (IBP) (OPTIONAL)
1 or 2 channels of invasive blood pressure, according to customer's choice.
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
• Range: -30 to 300 mmHg.
• Mode selection for Pressure Ranges: Automatic or Manual.
• Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
• Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
• Measure unit: mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

TEMPERATURE (OPTIONAL)
2 channels of temperature.
Compatible with Thermistor Series YSI 400.
Measurement of body temperature or injectate temperature during measurement of Cardiac Output, from -5ºC to 45ºC.
• Resolution: 0.1ºC.
• Accuracy: ±0.2ºC.
• Temperature Alarm, minimum and maximum: -5ºC to 45ºC and off.

Accessories included:
Feas Electrónica temperature sensor - DB9M / G.
Derivation box for Temperature 1 and Temperature 2 parameters, DB9M connector (for two Temperature channels).

CARDIAC OUTPUT (OPTIONAL)
• Entry of patient parameters: Weight, height.
• Selection of:
- Injection volume: 3, 5 and 10 ml.
- Catheter Constant K or Computing Constant (CC)
- Measurement of the temperature in blood and injected.
• Calculation of:
- Body surface.
- CO: Volume in liters / min.
- SV: Stroke volume.
- CI: Cardiac index.
- SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement when injecting.

Accessories included:
Catheter adapter cable for cardiac output - DB9M/G
Derivation box for parameters Temperature 1 and Cardiac Output with DB9.
Derivative box for parameters Temperature 1, Temperature 2 and Cardiac Output with DB9.

SIDESTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 170 1 / min.
• Accuracy:
- CO2, ± 2 mmHg for CO2 <5%.
- <10% of reading for CO2> 5%.
• Resolution:
- CO2, 1 mmHg.
- Respiratory Rate, 1 1 / min.
• Alarm:
- FiCO2, EtCO2, minimum and maximum, 0 to 100 mmHg and off.
- Minimum and maximum Respiratory Rate, 0 to 170 1 / min and off.
Calibration: 2 zero points (ambient air) and gain (5% or 10%).
Compensation for temperature and atmospheric pressure, O2 and N2O.

Accessories included:
Type "T" nozzle for Feás Electrónica lateral flow capnography.
2.4 m patient line for Feás Electrónica lateral flow capnography
Moisture filter for lateral flow capnography, Feas Electrónica

SIDESTREAM CAPNOGRAPHY WITH ANESTHETIC GASES MEASUREMENT (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1/min.
• Accuracy: CO2, ±3%ABS.
• Resolution: CO2, 1 mmHg.
- Respiratory Rate, 1 1/min.
• Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
• Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
• Measurement of N2 O: from 0% to 100%.
•Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate:
- 70 to 120 ml/min for neonatal patients.
- 120 to 250 ml/min for adult patients.

Accessories included:
"T" type nozzle for sidestream Capnography, Feas Electrónica.
2,4 m patient line for sidestream Capnography, Feas Electrónica
Water trap for use in lateral flow capnography and anesthetic gases.

MAINSTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1 / min.
• Accuracy: ± 2 mmHg between 0 and 40 mmHg and ± 5% of reading 41 and 70 mmHg.
• Resolution: 1 mmHg.
• Calibration: only zero (ambient air).
Compensation for temperature and atmospheric pressure, O2 and N2O.
Warm-up time <15s, to start measuring, within 2 minutes and obtain measurement values according to specifications.
• Waveform response time: less than 60 ms.

Accessories included:
Mainstream Capnography sensor, Feas Electrónica
Nozzle for mainstream Capnography sensor

FiO2 (OPTIONAL)
Fraction of inspired Oxygen: from 0% to 100%.
Polarographic cell measurement principle.
Alarms minimum and maximum: 0% to 100% and off.

Accessories included:
Derivative box for pressure 1, pressure 2 and / or FiO2 parameters with DB9M.
FiO2 Sensor, Feas Electrónica.
Disposable Oxygen Sensor T-adapter for FiO2, Feas Electrónica
Spiral cable of Oxygen Sensor for FiO2, Feas Electrónica.

ANY OTHER OPTIONAL CHANNEL SPECIFIED CAN BE ADDED IN THE FUTURE.

TECHNICHAL SPECIFICATIONS
15” LCD-TFT Color Monitor
• Resolution: 1024x768.
Wide angle of visibility.
• Dimensions:
- Height: 345mm; Width: 335mm; Depth: 205mm.
• Weight: 6Kg.
• Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
• Power Consumption: 85VA.

CERTIFICATIONS
• Quality Standard (Certificates):
- ISO 13485.
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

• Safety Standard (Certificates):
- UNE-EN60601-1.(Eq. IRAM 4220-1 and IEC 601-1).
- Electrical-shock protection: Class I – CF.
- UNE-EN60601-1-2 (Electromagnetic compatibility).
Authorized by A.N.M.A.T.: PM 1125-2.

ACCESSORIES
Included:3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
SpO2 extension cable.
Cuff for NIBP.
Silicone hose for NIBP.
Temperature sensor Feas Electrónica - DB9M/G.
Derivation box for parameters of Temperature 1 and Temperature 2, DB9M connector.
FiO2 Sensor Feas Electrónica.
T Adapter for FiO2 sensor Feas Electrónica.
Spiral cable for FiO2 sensor Feas Electrónica.
Power cord.
User manual.
The image is illustrative.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Non-invasive pressure cuff (18-26cm), reusable, pediatric, 1-tube, Feas Electrónica
One-way resusable cuff for non-invasive blood pressure measurement.
Size: 18 to 26 cm arm circumference.
With hose connector included.
For a small pediatric patient.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monitoring central station model 5032C Feas Electrónica
The Monitoring central station model 5032C Feas Electronica is a centralized patient monitoring system, composed of a Central Processing Unit (CPU) with CENTRALC software, a color LCD monitor, a keyboard, a mouse, a printer and the necessary wiring for the communication with bedside devices. The Monitoring Central allows monitoring from 4 to 16 patients, through the connection with bedside monitors. Monitors supported include, but are not limited to, the following devices: LCD Multipair Patient Monitor, Defibrillator Monitor 3850B, and FEAS ELECTRONICA Vital Signs Monitors. The connection between the Bedside Monitor and the Monitoring Center allows communication which can be: Analog, Digital and Full Digital.

- Provides simultaneous monitoring of electrocardiogram with indication of heart rate, respiratory curve with indication of respiratory rate and invasive pressure curve, with indication of systolic, diastolic and mean pressure values of up to 16 different patients.
- Ideal for use in intensive care and coronary units.
- Observation of Waveforms: electrocardiogram, invasive blood pressure and respiratory.
- Digital communicaton with bedside feas ELECTRONICA monitors, for the reception of values of: ECG, Respiratory, SpO2, Pulse Rate, EtCO2, Invasive Blood Pressure, Temperature, Non-Invasive Blood Pressure, Fico, and Respiratory Rate.
- Selection of ECG and RESPIRATORY curve gains.
- Auto scale invasive pressure curve.
- Curve sweep speed selection.
- Audible and visual alarms by maximum and minimum, configurable for all the parameters; with volume control.
- Enter text on screen (Patient Case History).
- Freeze waveform of selected patient.
- Observe previous events.
- Trend graphs. Graphic or tabular trend of the last 72 hours, displayed on 1, 4, 8, 12, 24, 48 and 72 hours.
- High resolution printing of: saved waveforms and values for heart rate, systolic, diastolic and mean pressures, respiratory rate and trends for all the parameters.

TECHNICAL SPECIFICATIONS

Analogic Input Signals:
- ECG: 1V/mV.
- PRESSURES: 1V/100 mmHg.
- RESPIRATORY: 1V/3 Ohm.

Digital connection to bedside monitors:
- Display: Color SVGA Monitor Dimensions: 19" (Optional large size).
- CPU: Industrial PC Motherboard. Storage Solid State Disk.
- Output for connecting additional remote monitors.
- Printer: Laser black and white.
- AC Voltge input: 90V-/240V. ~ Frequency 50Hz/60Hz.

ACCESSORIES INCLUDED
- Keyboard, mouse and pad.
- Power cords for CPU, monitor and printer.
- User manual.

AREAS OF USE:
Intensive and intermediate care units, coronary care units, operating rooms, rehabilitation centers, others.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Patient Monitor, Multiparameter, LCD, Feas Electrónica
Feas Electrónica MULTIPAR LCD Patient Monitor is designed for monitoring the vital signs of patients in Intensive Therapies, Coronary Units, Operating Rooms, Emergency Rooms, and in a wide variety of clinical-hospital environments, where several vital signs need to be monitored of a patient simultaneously.

GENERAL CHARACTERISTICS
FUNCTIONS:
• Menu, messages and instructions in English.
• Single sweep speeds for circulatory and respiratory system: 12.5, 25 and 50 mm/s.
• Visual and audible QRS indicator proceeding from ECG or Oxymeter.
• When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
• Three alarm levels: physiologicals, technicals and generals.
• Upper and lower limits adjustable by the user.
• Visual and audible alarm indicators, on screen and on top of the equipment.
• Hotkey to activate the silent alarms for 2 minutes.
• Alarms Turn Off.
• User friendly menu by using the selector knob and hotkeys.
• Waveform print-out on standard printer.
• Analog outputs for ECG, RESP, IBP and a synchronization pulse.
• Networking Connection for Centralized Monitoring.
• WAIT Mode: stops monitoring, which allows changing to another patient without activating the alarms.
• Screen menu: normal, 7 leads of ECG, Tabular trends or graphical trends, frozen and measuring of calipers, large numbers.

BASIC CHANNELS
ELECTROCARDIOGRAM
• Heart rate: 20 to 300 1/min.
• Gain selection: 1/4, 1/2, 1, 2 and 4 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
• 3 leads with 3 wire patient cable. Lead selection: DI, DII or DIII.
12 leads with 5 wire patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C1, C2, C3, C4, C5 or C6.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
• Heart rate alarms, minimum and maximum: 20 to 300 1/min and off.
Lead off detection.
Pacemaker pulse detection.
• Pacemaker pulse rejection in counting the heart rate:
•Pulse rejected characteristic:
- Amplitude: from ±2 mV to ±700 mV
- pulse width: from 0.1 ms to 2.0 ms.

ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
• Detects and classifies the following arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy, Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
• Store up to 32 events and allows review and printing.
• Measurement of the S-T segment: both the isoelectric point and the ST point are user adjustable (J-point +60 ms or +80 ms, or any other value in ms, from 1 to 120 ms).
• Range: ±2.00 mV.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
• SpO2:
- Range: 35% to 100%
- Precision:
±2% from 80% to 100%.
±3% from 70% to 79%.
- Resolution: 1%.
- Alarm: 0% to 100% and off.
- Initial alarm: 85%.
• Pulse rate:
- Range: 25 to 250 1/min.
- Precision: ±2 1/min.
- Resolution: 1 1/min.
- Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

Accessories included:
Finger pulse oximeter sensor.

RESPIRATORY WAVEFORM
Impedanciometric method.
• Measurement selection: Transabdominal or Transtoraxic.
• Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
• Respiratory rate alarm, minimum and maximum: 5 to 170 1/min and off.

NON-INVASIVE BLOOD PRESSURE (NIBP)
Oscilometric method.
• Use: ADULT or NEONATE.
• Measurement mode: AUTOMATIC or MANUAL.
• Measure unit: mmHg.
• Resolution:
- Pressure: 1 mmHg.
- Pulse rate: 1 1/min.
• Transductor accuracy: ±3 mmHg.
• Ranges:
- PEDIATRIC/ADULT
Systolic pressure: 25 to 265 mmHg.
Diastolic pressure: 12 to 220 mmHg.
Mean pressure: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.
- NEONATE
Systolic pressure: 25 to 135 mmHg.
Diastolic pressure: 12 to 110 mmHg.
Mean pressure: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.
• Alarm limits of:
- Systolic pressure from 20 to 255 mmHg and off.
- Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
• Protections:Double protection from over-pressure.
- Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
- Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
- Protection for exceeding time of measurement.

Accessories included:
Cuff for Non Invasive Blood Pressure.
Hose for Non Invasive Blood Pressure.


INVASIVE BLOOD PRESSURE (IBP) (OPTIONAL)
1 or 2 channels of invasive blood pressure, according to customer's choice.
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
• Range: -30 to 300 mmHg.
• Mode selection for Pressure Ranges: Automatic or Manual.
• Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
• Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
• Measure unit: mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

TEMPERATURE (OPTIONAL)
2 channels of temperature.
Compatible with Thermistor Series YSI 400.
Measurement of body temperature or injectate temperature during measurement of Cardiac Output, from -5ºC to 45ºC.
• Resolution: 0.1ºC.
• Accuracy: ±0.2ºC.
• Temperature Alarm, minimum and maximum: -5ºC to 45ºC and off.

Accessories included:
Feas Electrónica temperature sensor - DB9M / G.
Derivation box for Temperature 1 and Temperature 2 parameters, DB9M connector (for two Temperature channels).

CARDIAC OUTPUT (OPTIONAL)
• Entry of patient parameters: Weight, height.
• Selection of:
- Injection volume: 3, 5 and 10 ml.
- Catheter Constant K or Computing Constant (CC)
- Measurement of the temperature in blood and injected.
• Calculation of:
- Body surface.
- CO: Volume in liters / min.
- SV: Stroke volume.
- CI: Cardiac index.
- SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement when injecting.

Accessories included:
Catheter adapter cable for cardiac output - DB9M/G
Derivation box for parameters Temperature 1 and Cardiac Output with DB9.
Derivative box for parameters Temperature 1, Temperature 2 and Cardiac Output with DB9.

SIDESTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 170 1 / min.
• Accuracy:
- CO2, ± 2 mmHg for CO2 <5%.
- <10% of reading for CO2> 5%.
• Resolution:
- CO2, 1 mmHg.
- Respiratory Rate, 1 1 / min.
• Alarm:
- FiCO2, EtCO2, minimum and maximum, 0 to 100 mmHg and off.
- Minimum and maximum Respiratory Rate, 0 to 170 1 / min and off.
Calibration: 2 zero points (ambient air) and gain (5% or 10%).
Compensation for temperature and atmospheric pressure, O2 and N2O.

Accessories included:
Type "T" nozzle for Feás Electrónica lateral flow capnography.
2.4 m patient line for Feás Electrónica lateral flow capnography
Moisture filter for lateral flow capnography, Feas Electrónica

SIDESTREAM CAPNOGRAPHY WITH ANESTHETIC GASES MEASUREMENT (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1/min.
• Accuracy: CO2, ±3%ABS.
• Resolution: CO2, 1 mmHg.
- Respiratory Rate, 1 1/min.
• Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
• Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
• Measurement of N2 O: from 0% to 100%.
•Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate:
- 70 to 120 ml/min for neonatal patients.
- 120 to 250 ml/min for adult patients.

Accessories included:
"T" type nozzle for sidestream Capnography, Feas Electrónica.
2,4 m patient line for sidestream Capnography, Feas Electrónica
Water trap for use in lateral flow capnography and anesthetic gases.

MAINSTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1 / min.
• Accuracy: ± 2 mmHg between 0 and 40 mmHg and ± 5% of reading 41 and 70 mmHg.
• Resolution: 1 mmHg.
• Calibration: only zero (ambient air).
Compensation for temperature and atmospheric pressure, O2 and N2O.
Warm-up time <15s, to start measuring, within 2 minutes and obtain measurement values according to specifications.
• Waveform response time: less than 60 ms.

Accessories included:
Mainstream Capnography sensor, Feas Electrónica
Nozzle for mainstream Capnography sensor

FiO2 (OPTIONAL)
Fraction of inspired Oxygen: from 0% to 100%.
Polarographic cell measurement principle.
Alarms minimum and maximum: 0% to 100% and off.

Accessories included:
Derivative box for pressure 1, pressure 2 and / or FiO2 parameters with DB9M.
FiO2 Sensor, Feas Electrónica.
Disposable Oxygen Sensor T-adapter for FiO2, Feas Electrónica
Spiral cable of Oxygen Sensor for FiO2, Feas Electrónica.

ANY OTHER OPTIONAL CHANNEL SPECIFIED CAN BE ADDED IN THE FUTURE.

TECHNICHAL SPECIFICATIONS
15” LCD-TFT Color Monitor
• Resolution: 1024x768.
Wide angle of visibility.
• Dimensions:
- Height: 345mm; Width: 335mm; Depth: 205mm.
• Weight: 6Kg.
• Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
• Power Consumption: 85VA.

CERTIFICATIONS
• Quality Standard (Certificates):
- ISO 13485.
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

• Safety Standard (Certificates):
- UNE-EN60601-1.(Eq. IRAM 4220-1 and IEC 601-1).
- Electrical-shock protection: Class I – CF.
- UNE-EN60601-1-2 (Electromagnetic compatibility).
Authorized by A.N.M.A.T.: PM 1125-2.

ACCESSORIES
Included:3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
SpO2 extension cable.
Cuff for NIBP.
Silicone hose for NIBP.
Temperature sensor Feas Electrónica - DB9M/G.
Derivation box for parameters of Temperature 1 and Temperature 2, DB9M connector.
FiO2 Sensor Feas Electrónica.
T Adapter for FiO2 sensor Feas Electrónica.
Spiral cable for FiO2 sensor Feas Electrónica.
Power cord.
User manual.
The image is illustrative.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Pediatric pulse oximeter with Clamp and DB9M/G, Feas Electrónica 3178
Pediatric pulse oximeter with Clamp and DB9M/G, Feas Electrónica 3178

The reusable finger sensor is used in oximetry monitoring systems to non-invasively determine Oxygen saturation (SpO2), pulse rate, and pulse plethysmographic waves.

The reusable pediatric finger sensor is recommended for:
• Patients weighing more than 5Kg and less than 45Kg.
• Patients with limited activity.
• For sampling.

The reusable pediatric finger sensor is not recommended for long-term use. For long-term patient monitoring, use disposable single-patient or wrap-around type sensors.
It has a length of 1 meter.
Protection against liquids IPX1.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Adult Oximeter Finger Sensor DB9M/G Feas Electrónica
Adult Oximeter Finger Sensor DB9M/G Feas Electrónica

The Oximeter sensor is applicable to adult patients (weighing more than 30kg), for non-invasive measurement by continuous sampling of arterial oxygen saturation and pulse rate, in Oximetry Monitoring Systems. This sensor is designed to be used exclusively in Feas Electronica equipment.

It has a length of 2.5 meters.
Protection against liquids: IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Adult pulse oximeter sensor with clamp and DB9M/G 3444, Feas Electrónica
Adult pulse oximeter sensor with clamp and DB9M/G 3444, Feas Electrónica

The reusable finger sensor Feas ELECTRONICA ref. 5043 is used to non-invasively determine oxygen saturation (SpO2), pulse rate, and pulse plethysmographic waves. It is indicated for:
• Patients who weigh more than 15Kg.
• Patients with limited activity.

The reusable finger sensor is not recommended for prolonged use, unless the integrity of the skin is checked frequently and the application site is changed accordingly. For long-term patient monitoring, use either single-patient disposable sensors or Velcro-strap sensors.
It has a length of 1 meter.
Protection against liquids IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Neonatal oximeter sensor DB9M/G Feas Electrónica 3026
Neonatal oximeter sensor DB9M/G Feas Electronica 3026.

This sensor is used in oximetry monitoring systems to non-invasively determine Percent Oxygen Saturation (SpO2), pulse rate, and pulse plethysmographic waves.
Its use is indicated for patients of less than 3kg.

It has a length of 40 cm.
Protection against liquids IPX1.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Oxygen sensor with tongue clips for veterinary use. D9MG, Feas Electrónica. 551A
Oxygen sensor with tongue clips for veterinary use. D9MG, Feas Electronica.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Polycarbonate NIBP parameter, revision 01, front of Multipar LCD version 7.x
Self-adhesive polycarbonate for multipair monitor of the NIBP parameter connector.
Self-adhesive polycarbonate of "SILENCE" key for Multipar keyboard
Self-adhesive polycarbonate of "SILENCE" key for Multipar keyboard

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Self-adhesive polycarbonate of key "ESC" for keyboard of Multipar
Self-adhesive polycarbonate of key "ESC" for keyboard of Multipar

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Gray nylon KNOB FOR .250" electronic switch PUSHBUTTON
Grayhill trim knob. Replacement for rotary knob of the Multipair Monitor

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Patient Monitor, Multiparameter, Comen Model C30
The COMEN C30 Multi-Parameter Patient Monitor is the perfect device to meet all your portable monitoring needs in and out of the hospital. In addition, its hardness and quality allow it to withstand a 1.2 meter drop, making it an excellent portable device that meets the needs of all hospitals. Created for use in adult, pediatric and neonatal patients. It can provide users with all the basic monitoring parameters required by their patients.
Strong, stable, waterproof, dustproof, anti-drop and shockproof.
Small and light, easy to carry.
Suitable for emergency scene, out-of-hospital transport, intra-hospital transport, bedside monitoring in the emergency department

Model C-30
Technical specifications
Security
ISO 13485: 2003 Approved
CE marking according to MDD93 / 42 / EEC

Physical characteristics:
Dimensions: 190mmX81.6mmX107mm
weight: 1.2 kg

Operation Environment
Power: 15Vdc
Power supply: 100-250Vac, 50/60 Hz.
Temperature: 0-40 ° C
Humidity: 15-85%
Battery: Lithium-ion, rechargeable
Plug & play battery, capacity: 2600mA
Maximum of 6.5 hours for charging;
4 hours of continuous work

Specifications
Patient: adult, pediatric, neonate
Display: 4.3 "color TFT touch screen
Resolution: 16:9
2 curves on screen
Sweep speed: 12.5mm / s, 25mm / s, 50mm / s
Indicators:
• Battery indicator light
• Dual indicator alarm light
QRS beep and alarm sound
Key sound
AC input socket
MDR
Trend: 48 hours
Alarm: user adjustable high and low limits
Connection to the monitoring center
DVI connection

Breathing
Impedance method
Rate:
• Adult 0-1OOrpm
• Neonatal 0-150 rpm
• Resolution: ± 1 rpm
Apnea alarm: 10s-40s
Audible and visual alarm
Sweep speed: 12.5.25mm/s
Gain selection: x0.25, x0.5, x1, x2

ECG
Cable: 12-lead CardioTecTM,
3 or 5 snaps (selectable)
12 leads: I, Il, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, -V6
5 wire: RA, LA, RL, LL, V or R, L, N, F, C,
3 wire: RA, LA, LL or R, L, F
Lead Selection: 12 leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
5 wire: I, Il, IIl, aVR, aVL, aVF, V
3 wire: I, Il, III
Gain selection: x0.25, x0.5, x1, x2, auto
Adult: 15-300 1 / min
Pediatric / neonate: 15-350 1 / min
Accuracy: ± 1% or ± 1 1 / min (whichever is greater)
Resolution: 1 1 / min
ECG bandwidth:
MON: 0.5-55Hz
DIA: 0.05-150 Hz
OPE: 1-20Hz
ST segment detection: ± 2.0mV (automatic)
Arrhythmia Analysis
Pacemaker pulse detection

Temperature
2 channels
Standard sensor: Cutaneous
Measurement range: 0-50 ° C
Resolution: 0.1 ° C
Accuracy: 0.1 ° C (excludes sensor error)
Channel: T1, T2, TD (temperature difference)

PNI
Measurement method: Automatic oscillometric
Working mode: manual / automatic / continuous
Interval: Adjustable from 1 min to 480 min
Measurement types: systolic, diastolic, mean
Unit of measure: mmHg / KPa, adjustable
Systolic pressure range:
• Adult: 40 to 270 mmHg
• Pediatric: 40 to 200mmHg
• Neonate: 40 to 135 mmHg
Diastolic pressure range:
• Adult: 10 to 215 mmHg
• Pediatric: 10 to 150mmHg
• Neonate: 10 to 100 mmHg
Medium pressure range:
• Adult: 20 to 235 mmHg
• Pediatric: 20 to 165mmHg
• Neonate: 20 to 110 mmHg
Overpressure Protection: Hardware and Software
• Accuracy: less than ± 5mmHg
• Resolution: 1mmHg

Sp02 Comen
Range: 1-100%
Resolution: 1%
Accuracy:
• ± 2% (70-100%, Ped, no movement)
• ± 3% (70-100%, Neo, no movement)
• 0-69% unspecified
Range alarm: 0-100%
Pulse Frequency:
• Range: 20 to 300 ppm
• Resolution: 1 ppm
• Accuracy:
- ± 3ppm (no movement)
- ± 5ppm (movement)
Alarm: 20 to 300 ppm

EtCO2
Sidestream CO2 (SideStream)
Range: 0 to 150 mmHg, 0 to 79%,
• 0 to 20 kPa (at 760 mmHg)
• Accuracy: ± 2mm Hg (0-40mmHg)
• ± 5% of reading (41-70 mmHg)
• ± 8% of reading (71-100 mmHg)
• ± 10% of reading (101-150 mmHg)
Frequency: 0-120 rpm
Accuracy:
• ± 2rpm (0-70rpm)
• ± 5 rpm (> 70 rpm)
Response time: <240 ms (10% to 90%)
Delay time: <2s

C02 COMEN Main Flow
Range: 0 to 150 mm Hg, 0 to 79%,
0 to 20 kPa (at 760 mmHg)

Accuracy: ± 2mmHg (0-40mmHg)
• ± 5% of reading (41-70 mmHg)
• ± 8% of reading (71-100 mmHg)
• ± 10% of reading (101-150 mmHg)
Frequency: 0-120 rpm
Accuracy:
• ± 2rpm (0-70rpm)
• ± 5 rpm (> 70 rpm)
Delay time: <2s

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
AC inlet EMI/RFI filter 04SS3A-2SR-Q fused 4A-250VAC RoHS H&L corp-PN: 530040302014100
EMI/RFI filter 04SS3A-2SR-Q, 4A 250VAC-PN: 530040302014100

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Disk on module (DOM), 128MB IDE EDC 4000 40 pin vertical
The Innodisk vertical EDC 4000 adapts to any platform with an IDE connector and provides a quick and easy path upgrade. Designed for embedded industrial applications that require a quick boot process, the EDC 4000 efficiently updates the work environment and improves data transfer rates with read and write performances of up to 40MB and 20MB per second, respectively. The vertical EDC 4000 has 128MB capacity and 40-pin vertical connector packages.

The integrated disk card is a replacement for the central monitoring station.

Months Warranty: 3

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Rotary opto encoder 16 position (trim knob)
16-position rotary optical encoder or rotary knob for menu access on monitors Mod. Multipar from Feas Electrónica.
High reliability, with push button for Enter key or data confirmation16-position rotary optical encoder or rotary knob for menu access on monitors Mod. Multipar from Feas Electrónica.
High reliability, with push button for Enter key or data confirmation


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.