MULTIPARAMETER PATIENT MONITOR Model: PM9000
GENERAL CHARACTERISTICS
Features:
Menu, messages and instructions in English or Spanish.
Single sweep speeds for circulatory (25mm/s y 50mm/s) and respiratory system (6,25 mm/s, 12,5 mm/s, 25 mm/s y 50 mm/s).
Visual and audible QRS indicator proceeding from ECG or Oxymeter.
When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
Physiological and technical alarms.
Three alarm priority levels: low, medium and high priority.
Upper and lower limits adjustable by the user.
Visual and audible alarm indicators, on screen and on top of the equipment.
Hotkey to pause visual or audible alarms for 2 minutes.
Visual or audible alarms Turn Off function.
User friendly menu by using the trim knob and hotkeys.
Networking Connection for Centralized Monitoring.
Screen menu: - Normal.
- 7 leads of ECG.
- Tabular trends or graphical trends.
- Big numbers
- OxyCRG
BASIC CHANNELS
Electrocardiogram (ECG)
Heart rate: Adult 15 to 300 1/min. Pediatric 15 a 350 1/min
Gain selection: 0.25, 0.5, 1 and 2 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
3 leads with 3 leads patient cable. Lead selection: DI, DII or DIII.
7 leads with 5 leads patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
Heart rate alarms: Lower and upper (medium priority).
Lower and upper extreme (high priority) and off.
Lead off detection.
Pacemaker pulse indication.
Pacemaker pulse rejection in counting the heart rate:
Pulse rejected characteristic: -Amplitude: from ±2 mV to ±700 mV.
-Pulse width: from 0.1 ms to 2.0 ms.
Analysis and classification of arrhythmias
Detected and classified the follow arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy,
Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
Measurement of the S-T segment: both the isoelectric point and the ST point areuser adjustable.
Range: ±2.00 mV. PVCs per minute measurement.

Respiratory waveform (RESP)
Impedancemetry Method.
Measurement selection: Transabdominal (II-F) o Transthoracic (I-L).
Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
Respiratory rate alarm, minimum and maximum: 5 to 170 1/min.

Oxygen saturation (SpO2)
Plethysmographic waveform, Display of percentual oxygen saturation.
Measurement signal low perfusion and motion.
SpO2:
Range: 0% to 100%.
Adult precision: 2% from 70% to 100%.
Neonate precision: 3% from 70% to 100%.
Motion precision: 3% from 70% to 100%.
Resolution: 1%.
Alarm: 0% to 100% and off.
Initial alarm: 85%.
Pulse rate:
Range: 20 to 250 1/min.
Adult precision: ±3 1/min.
Neonate precision: ±3 1/min.
Motion precision: ±5 1/min.
Resolution: 1 1/min.
Alarm: 20 to 250 1/min and off.

Two channels of invasive blood pressure (IBP)
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
Range: -30 to 300 mmHg.
Mode selection for Pressure Ranges: Automatic or Manual.
Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
Technical alarm due to transducer failure and disconnected catheter.
Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
Units in mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

Two temperature channels
Compatible with Thermistor Series YSI 400.
Measurement of body temperature (25°C a 50°C) or injectate temperature during measurement of Cardiac Output, from (-10ºC to 25ºC).
Resolution: 0.1ºC. Accuracy: ±0.2ºC.
Temperature Alarm, minimum and maximum:(25ºC to 50ºC) and off.

Cardiac output (CO)
Input of patient parameters: Weight, height.
Selection of: Injectate volume: 3, 5 and 10 ml.
Catheter Constant K or Computing Constant (CC).
Measurement of temperature in blood and injected.
Calculation of: Body surface.
CO: Volume in liters/min.
SV: Systolic volume.
CI: Cardiac index.
SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement, at perform the injection.
Stores 8 measurements.
Blood temperature: Range 15°C to 45°C, resolution 0.01°C and accuracy ±0.3°C.
Injected temperature: Range -10°C to 25°C, resolution 0.1°C and accuracy ±0.4°C.

Non-invasive blood pressure (NIBP)
Oscillometric method
Use: ADULT/PEDIATRIC or NEONATE.
Measurement mode: AUTOMATIC or MANUAL.
Unit in mmHg.
Resolution:
Pressure: 1 mmHg.
Pulse rate: 1 1/min.
Transductor accuracy: ±3 mmHg or ±2%, whichever is greater.
Ranges:
ADULT/PEDIATRIC
Pressure: Systolic: 25 to 265 mmHg.
Diastolic: 12 to 220 mmHg.
Mean: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.

NEONATE
Pressure: Systolic: 25 to 135 mmHg.
Diastolic: 12 to 110 mmHg.
Mean: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.

Alarm limits of: Systolic pressure from 20 to 255 mmHg and off.
Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
Protections: Double protection from over-pressure.
Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
Protection for exceeding the time of measurement.

OPTIONALS
FiO2
Fraction of inspired Oxygen: from 0% to 100%.
Galvanic cell measurement principle.
Alarms minimum and maximum: 18% to 100% and off.

Recorder
Direct access key for record command.
High resolution thermal array printer.
Recorder speed: 25 mm/s.
Recorded in real time up to 2 selectable waveforms between ECG, RESP, IBP,
SpO2 and CO2 waveform.
Prints the header with date and time.
Print values of Sweep Speed, Lead, Gain, Respiratory Rate, Heart Rate,
Systolic, Diastolic and Mean Pressure, SpO2 and Pulse Rate, EtCO2, FiCO2
and Respiratory Rate (CO2), according recorded waveform.

Capnography

Option 1: Sidestream
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate: 4 to 150 1/min.
Accuracy: from 0 mmHg to 38 mmHg: ±2.0 mmHg
from 39 mmHg to 70 mmHg: ±5 mmHg
from 71 mmHg to 100 mmHg: ±8 mmHg
Resolution: CO2, 1mmHg.
Respiratory Rate, 1 1/min.
Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100mmHg and off.
Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
Calibration: Two zero points -room air and gains (5% or 10%)-.

Option 2: Sidestream with anesthetic gases measurement
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate, 4 to 150 1/min.
Accuracy: CO2, ±3%ABS.
Resolution: CO2, 1 mmHg.
Respiratory Rate, 1 1/min.
Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
Measurement of expired and inspired O2: from 0% to 100%, Resolution: 1%.
Measurement of N2O: from 0% to 100%.
Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,
Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate: 70 to 120 ml/min for neonatal patients.
120 to 250 ml/min for adult patients.

Option 3: Mainstream
Measurement of EtCO2, FiCO2 and Respiratory Rate.
Range: CO2, from 0 to 100 mmHg (0% to 13%).
Respiratory Rate, 4 to 170 1/min.
Accuracy: ±2 mmHg from 0 to 40 mmHg.
±5% of reading from 41 to 70 mmHg.
Resolution: 1 mmHg.
Calibration: only zero (room air).
Compensation for temperature and atmospheric pressure, O2 y N2O.
Warm-up time <15s, to start measuring, within 2 minutes to obtain values consistent with the specifications.
Waveform response time: less than 60ms.

Touchscreen

ACCESSORIES
Included:
3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
Pressure Transducer Adaptor (2 units).
Pressure transducer (2 units).
Temperature sensor.
Power cord.
User manual.

TECHNICAL SPECIFICATIONS
15” LCD Color Monitor. Resolution: 1024 x 768. Greater angle of view.
Dimensions: Height: 345mm; Width: 335mm; Depth: 205mm.Weight: 6 Kg.
Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
Power Consumption: 65VA.
Internal Battery: 14.4V / 8.8Ah Li-ion battery. Autonomy: 3.5 hours.
Battery state indication (empty, full, on charge) on screen.
Quality Standard: ISO 13485, Disp. A.N.M.A.T. 3266/2013: B.P.F. and
Disp. A.N.M.A.T. 2319/02, certified.
Safety Standard: UNE-EN60601-1. (Eq. IRAM 4220-1 and IEC 601-1) Electrical-
shock protection: Class I - CF, UNE-EN60601-1-2 (EMC), certified.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.

Documentation of the Patient Monitor model PM9000 Feas Electronica

- See user manual
- Download Certificate ANMAT