Monthly rent - Feas Electrónica's Automatic External Defibrillator (AED), model: Heart+ ResQ NT-381.C
GENERAL CHARACTERISTICS
Defibrillator
• Waveform: truncated exponential biphasic.
• 150 J (Adult, impedance 50 Ω).
• 50 J (Pediatric, impedance 50 Ω).
Charging time: 9 seconds (± 10 %).
• Shock method: manual button (flashes when ready).
• Mode conversion: Adult/Child toggle button.
(Child Mode delivers 50J Pediatric shock for patients weighing less than 25 Kg, approximately 12 months to 8 years old).

ECG
• ECG record: LEAD II.
• Analysis overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation.
• Analysis time: Approximately 7 seconds.

Battery Check
• Battery reserve level:
High: 3 bars.
Medium: 2 bars.
Low: 1 bar, red status led.
Critical: no bars, red status led.
• Battery status audible warnings.
Critical: Voice prompt states “Replace the battery”.

Data
• Stored data: ECG recordings, operation information (whether power is delivered or not), ambient noise, self diagnostic history.
• Memory capacity:
- ECG: Consecutively make records during analysis and CPR (once in every 5 s, up to 5000 s).
- Audio recording: 60 min in total (4 recording files of 15 min).
- Self-diagnosis: 3000 cases and results.
• Storage type: Internal memory.
• Access/search: “NT-ADV” software (for PC), via Bluethooth.

Audible and Visual Instructions
• After the equipment is switched on, the voice instructions will guide the operator through the resuscitation process, including the positioning of the patches, the indication of “Don’t touch the patient” (during the energy analysis and discharge process) and CPR (Cardio-Pulmonary Resuscitation) instructions.
• The CPR instructions tell the operator when to start Cardio-Pulmonary Resuscitation, according to the AHA Guide.

TECHNICAL SPECIFICATIONS
• Dimensions: 220 mm (W) x 282 mm (L) x 80 mm (H).
• Weight: 2 Kg (including battery and pads).
• Internal non-rechargeable Li-MnO2 battery - 12 Vdc/4200 mAh.
• Protection against ingress of liquids: IPX5 according to IEC60529.
• Temperature:
Storage: 15 ºC to 55 ºC.
Operation: 0 ºC to 45 ºC.
• Humidity:
Storage: 0 % to 95 % (without condensation).
Operation: 0 % to 90 % (without condensation).
• Altitude (Storage/Operation): 0 to 5000 meters.
• Drop test: 1.5 m according to IEC 60068-2-32.
• Quality standards:
ISO 13485 (certified).
Available A.N.M.A.T. 3266/2013: B.P.F.
Available A.N.M.A.T. 2319/02.
• Safety standards:
- IEC 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- IEC 60601-1-2 (Electromagnetic Compatibility).
- IEC 60601-1-6.
- IEC 60601-1-11.
- IEC 60601-2-4.

ACCESSORIES INCLUDED
- Electrode pads
• Shelf life: 2 (two) years.
• Single use.
• Cable length: 2.1 m (± 1 %).
• Applicable ages: All ages (12 months & older).
• Sealing: Airtight.
• Size: 125 mm x 95 mm.
• Pads placement:
Adult: anterior-lateral.
Pediatric: anterior-posterior (see positioning details on the package label).
- 1 Battery Pack
• Type: non-rechargeable Lithium-Manganese Dioxide (Li-MnO2).
• Voltage: 12 Vdc/4200 mAh.
• Standby mode life: Recommended five (5) years @ 25 ºC.
• Capacity: 180 shocks @ 25 ºC, 130 shocks @ 0 ºC.
• Measures: 22 mm (H) x 164 mm (W) x 54 mm (D).
• Weight: 260 g (±10 %).
- 1 User manual
- 1 bag
- Event Reading Software

OPTIONAL
USB/Bluetooth adapter.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Peristaltic infusion pump model SN-1500H
RENT
Includes equipment rental. For different rental periods, please consult.

Infusion pump of volumetric type with peristaltic perfusion system that accepts universal type set. Comprehensive and effective solution to optimize infusion therapy in all the units of the Hospitals and Health Centers.

• Great quality. Easy to use. Menu in Spanish.
• It has a simple system to calibrate different sets of infusion and can be recorded up to 9 calibrations for different sets.
• It has a horizontal structure with an effective prevention against the entry of liquids. It can be hung from the serum support or placed on a table or car.
• High accuracy, due to an intelligent pulsating compensation made by a double micro-processed unit, resulting in a constant flow of high precision infusion.
• It has five types of infusion mode: Speed ​​mode, Drip mode, Body Weight mode, Time mode and Drug mode.
• Contains drip sensor.
• Contains ultrasonic bubble sensor, high sensitivity.
• It contains a pressure sensor that allows adjustment of 3 levels of occlusion pressure.
• High definition screen, allows the on-screen display of the controlled / running function.
• Illuminated alarm panel (up to 12 alarms). Interface for nurse call.
• Record of 1500 historical events.
• The following features and functions contribute to make the patient's treatment more pleasant:
• Preheating function of the infused solution, with constant temperature of the solution.
• When the pump door is opened, in a dark environment, a light is switched on automatically that allows changing the set or working on the infusion pump without turning on the ambient light.
• Silent operation and without vibrations, because it is equipped with a high quality motor.
• External DC power supply input, also works with 12 VDC input.
• Easy to transport due to its small size and weight (2.5 kg).

TECHNICAL SPECIFICATIONS
Specification of IV set: 20 drops / ml or 60 drops / ml.

Infusion flow rate:
•Set IV of 20 drops / ml:
0.1ml / h ~ 1500ml / h.
1 drops / min ~ 350 drops / min.
• Set IV of 60 drops / ml:
0.1ml / h ~ 200ml / h.
1 drops / min ~ 200 drops / min.
• When the speed is between 0.1ml / h ~ 99.9ml / h, the infusion flow rate is changed from 0.1ml / h.
• When the speed is above 100ml / h, the infusion flow rate is changed from 1ml / h.
• Accuracy: ± 5%.

Bubble Detector:
• Ultrasonic detection method: bubbles are detected above 25μl.

- Occlusion warning speed:
It can be divided into three ranges as indicated,
- High (H) 900mmHg ± 200mmHg (120kPa ± 26,7kPa).
- Central (C) 500mmHg ± 100mmHg (66,7kPa ± 13,3kPa).
- Low (L) 100mmHg ± 50mmHg (13.3kPa ± 6,6kPa).
• Flow rate of the Purge Bolus:
Adjustable between 600ml / h ~ 1000ml / h.
• Kvo rate (to maintain permeable route):
Adjustable minimum flow between 0.1ml / h ~ 5ml / h.

Built-in battery:
• Ni-Mh, 12 Vdc, 2300 mAh, the recharge cycle can be repeated up to 500 times.
• Operating time of the battery: with the battery fully charged, it can run continuously for more than 5 hours at a speed of 25ml / h.

Operation mode:
Keep going.

Protection against the ingress of liquids: IPX4 - Protected against splashing water in any direction.

Historical records:
You can store the historical records of more than 1500 infusions made.

Communication link: SN-1500H: RS-232.

• Supply voltage: 110V ~ to 230V ~ - 50Hz or 60Hz or 12Vcc-5A external.
• Maximum power: 40 VA.
• Classification: class I and internally powered.
• Protection against electric shocks: Type CF.

WORK ENVIRONMENT:
Operating conditions:
• Ambient temperature: + 5ºC to + 40ºC.
• Relative humidity: 20% to 90%.
• Atmospheric pressure: 645mmHg (86kPa) to 795mmHg (106kPa).
Storage conditions:
• Ambient temperature: -5ºC to + 55ºC.
• Relative humidity: 0% to 95%.
• Environmental pressure: 500mmHg (66,7kPa) to 797mmHg (106,3kPa).

ACCESSORIES INCLUDED
1 Power cable to line.
1 user manual.

CERTIFICATES
• Certified Quality Standard:
ISO13485.
Disp. A.N.M.A.T. 3266/2013: B.P.F.
Disp. A.N.M.A.T. 2319/02.
• Safety rules:
IEC 60601-1 (Eq IRAM 4220-1 and IEC 601-1- Req.
IEC 60601-1-2 (Electromagnetic Compatibility).
IEC 60601-2-24 (Infusion Pumps).

FREQUENT QUESTIONS
Do you use any brand of infusion set?
Yes, the pump is not dependent on any brand of supplies. It is recommended that the silicone IV Set be used.

Can it be used on animals?
Yes, it can be used on animals.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Peristaltic infusion pump model SN-1800V
Includes equipment rental. For different rental periods, please consult.

Infusion pump of volumetric type with peristaltic perfusion system that accepts universal type set. Comprehensive and effective solution to optimize infusion therapy in all the units of the Hospitals and Health Centers.

• Great quality. Easy to use. Menu in English.
• It has a simple system to calibrate different sets of infusion and can be recorded up to 12 calibrations for different sets.
• It has a vertical structure with an effective prevention against the entry of liquids. It can be hung from the serum support or placed on a table or car.
• High accuracy, due to an intelligent pulsating compensation made by a double micro-processed unit, resulting in a constant flow of high precision infusion.
• It has six types of infusion mode: Speed mode, Drip mode, Body Weight mode, Time mod, Infusion Plan and Drug Library or Drug Mode.
• Contains drip sensor.
• Contains ultrasonic bubble sensor, high sensitivity.
• It contains a pressure sensor that allows adjustment of 10 levels of occlusion pressure.
• High definition screen, allows the on-screen display of the controlled / running function.
• Illuminated alarm panel (up to 12 alarms). Interface for nurse call.
• Record of 1500 historical events.

The following features and functions contribute to make the patient's treatment more pleasant:
• When the pump door is opened, in a dark environment, a light is switched on automatically that allows changing the set or working on the infusion pump without turning on the ambient light.
• Silent operation and without vibrations, because it is equipped with a high quality motor.
• External DC power supply input, also works with 12Vdc input.
• Easy to transport due to its small size and weight.
• Interface USB.

TECHNICAL SPECIFICATIONS
Specification of IV set: 20 drops/ml or 60 drops/ml.

Infusion flow rate:
• Set IV of 20 drops/ml: 0.1ml/h ~ 1500ml/h.
• Set IV of 60 drops/ml: 0.1ml/h ~ 1500ml/h.
• When the speed is between 1ml/h ~ 99.9ml/h, the infusion flow rate is changed from 0.1ml/h.
• When the speed is above 100ml/h, the infusion flow rate is changed from 1ml/h.
• Accuracy: ± 5%.

Bubble Detector:
Ultrasonic detection method: bubbles are detected above 25μl.

Occlusion pressure value:
• Ten adjustable levels, with an error of ± 50mmHg or ± 25% (6.6KPa).
• Adjustment range: from 100mmHg (13.3KPa) to 900mmHg (120KPa).
Flow rate of the Purge Bolus:
• Adjustable between 5ml/h ~ 1000ml/h.
Kvo rate (to maintain permeable route):
• Adjustable minimum flow between 1ml/h ~ 5ml/h.

Built-in battery:
• Ni-Mh, 12 Vdc, 2300 mAh, the recharge cycle can be repeated up to 500 times.
• Operating time of the battery: with the battery fully charged, it can run continuously for more than 6 hours at a speed of 5ml/h.

Operation mode: Continuous.

Protection against the ingress of liquids: IP32.

Historical records:
You can store the historical records of more than 1500 infusions made.

Communication link: SN-1800V: RS-232.

• Supply voltage: 100V ~ to 240V ~ / 50Hz or 60Hz or 12Vdc external.
• Maximum power: 28 VA.
• Classification: class I and internally powered.
• Protection against electric shocks: Type CF.

Work environment:
• Operating conditions:
Ambient temperature: + 5ºC to + 40ºC.
Relative humidity: 20% to 90%.
Atmospheric pressure: 645mmHg (86kPa) to 795mmHg (106kPa).
• Storage conditions:
Ambient temperature: -20ºC to + 55ºC.
Relative humidity: 0% to 95%.
Environmental pressure: 375mmHg (50kPa) to 795mmHg (106kPa).

Weight: 1,8Kg.

Dimensions: 129mm x 130mm x 215mm (Width x Length x Height).

ACCESSORIES INCLUDED
1 Power cable to line.
1 user manual.

CERTIFICATES
• Certified Quality Standard:
ISO 13485.
Disp. A.N.M.A.T. 3266/2013: B.P.F.
Disp. A.N.M.A.T. 2319/02.
• Safety Standard:
IEC 60601-1 (Eq IRAM 4220-1 and IEC 601-1- Req.
IEC 60601-1-2 (Electromagnetic Compatibility).
IEC 60601-2-24 (Infusion Pumps).

FREQUENT QUESTIONS
Do you use any brand of infusion set?
Yes, the pump is not dependent on any brand of supplies. It is recommended that a silicone IV Set be used.

Can it be used in animals?
Yes, it can be used in animals.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - One-channel syringe infusion pump
RENT
Includes equipment rental. For different rental periods, please consult.

Syringe infusion pump model SN-50C66T is a constant volume, channel and speed controlled infusion pump. It is characterized by high precision, stable flow rate and low dosage.

It is used particularly for the infusion of sodium nitroprusside, dopamine, propofol and antibiotics for acute and serious illnesses. The pump has multiple types of alarm functions to ensure a safe and reliable infusion process.

This product is used to clinically control the flow and velocity of fluids (drugs, nutrients, and blood) infused into the patient's body for safer use.

GENERAL CHARACTERISTICS
* The equipment can store customized syringe information from 33 different manufacturers, Weigao and 32 other manufacturers, configurable by the user.
* SN-50C66T pump can use sterile and disposable syringes of 5ml, 10ml, 20ml, 30ml or 50ml. Automatically recognizes the syringe in place.
* The equipment can store up to 2000 historical infusion records, which include the following information: injection mode, injection rate, alarm information, total volume injected, threshold value of the occlusion pressure limit of the path, limiting quantity, number of syringe and information of the operation.
* It has the following alarm functions:
Incorrect position of the syringe.
Residual limit.
Occlusion.
End of injection.
No operation.
From empty syringe.
Flow limit exceeded.
Plunger decoupled.
Power cord disconnected.
Discharged battery.
Dead battery.
System error.
* Battery operating time: starting from a full charge, the battery can support 6 hours of operation on one channel at a rate of 5 ml / h.
* Allows operation and battery charging from the car's 12Vdc connector.
* Intended User: Adult, Pediatric, Neonatal

TECHNICAL SPECIFICATIONS
* Dimensions:
306mm x 140mm x 146mm (Length x Width x Height)
* Weight: 2.25kg (includes support).
* Electrical feeding:
AC input: 100V to 240V, 50/60 Hz.
DC input: 12Vdc.
Battery voltage: 12Vdc.
Nominal power: 40 VA.

CERTIFICATES:
* Quality standards:
ISO13485
Disp. A.N.M.A.T. 3266/2013: B.P.F.
Disp. A.N.M.A.T. 2319/02
* Safety rules:
UNE-EN 60601-1. (Eq. IRAM 4220-1 and IEC 601-1), Class I - Prot. Desc. Elect .: CF
UNE-EN 60601-1-2 (Electromagnetic Compatibility).
Authorized by the A.N.M.A.T .: PM 1125-18.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Two-channel syringe infusion pump
RENT
Include equipment rental. For different rental periods, please consult

Syringe infusion pump model SN-50F66 is a constant volume, channel and speed controlled infusion pump. It is characterized by high precision, stable flow rate and low dosage.

It is used particularly for the infusion of sodium nitroprusside, dopamine, propofol and antibiotics for acute and serious illnesses. The pump has multiple types of alarm functions to ensure a safe and reliable infusion process.

This product is used to clinically control the flow and velocity of fluids (drugs, nutrients, and blood) infused into the patient's body for safer use.

GENERAL CHARACTERISTICS
* The equipment can store customized syringe information from 33 different manufacturers, Weigao and 32 other manufacturers, configurable by the user.
* SN-50 pumps can use sterile and disposable syringes of 5ml, 10ml, 20ml, 30ml or 50ml, depending on the model. Automatically recognizes the syringe placed.
* The equipment can store up to 2000 historical infusion records, which include the following information: injection mode, injection speed, alarm information, total injected volume, threshold value of the occlusion pressure limit of the path, limiting quantity, number of syringe and information of the operation.
* It has the following alarm functions:
Incorrect position of the syringe.
Of residual limit.
Occlusion
End of injection.
Without operation
Empty syringe
Flow limit exceeded.
Plunger decoupled.
Power cable disconnected.
Of discharged battery.
Battery exhausted.
Of system error.
* Operating time of the battery: starting from a full charge, the battery can support 6 hours of operation in a channel at a speed of 5 ml/h and 4 hours of operation in two channels.
* Allows battery operation and charging from the car's 12Vdc connector.

TECHNICAL SPECIFICATIONS
* Dimensions:
354mm x 195mm x 148mm (Length x Width x Height)
* Weight: 3.45kg (includes support).
* Electrical feeding:
AC input: 100V ~ to 240V ~, 50/60 Hz.
DC input: 12Vdc.
Battery voltage: 12Vdc.
Nominal power: 40 VA.

CERTIFICATES
* Quality Standards:
ISO13485
Disp. A.N.M.A.T. 3266/2013: B.P.F
Disp. A.N.M.A.T. 2319/02; certified
* Safety Standards:
UNE-EN 60601-1 (Eq IRAM 4220-1 and IEC 601-1), Class I - Prot. Desc. Elect .: CF,
UNE-EN 60601-1-2 (Electromagnetic Compatibility).
Authorized by the A.N.M.A.T .: PM 1125-18.

ACCESSORIES
● 1 Power cable.
● 1 User's manual.
● 1 Content list.
● 1 Manufacturing quality certificate.
● 1 Warranty card.
The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Three-channel syringe infusion pump
GENERAL CHARACTERISTICS
* It is characterized by its high precision, stable flow rate and low dosage. It is particularly used for the infusion of sodium nitroprusside, dopamine, propofol and antibiotics for acute and severe diseases.
* The equipment can store personalized syringe information from 33 different manufacturers, Weigao and 32 other manufacturers, configurable by the user.
* The SN-50F66 pump can use sterile and disposable syringes of 5ml, 10ml, 20ml, 30ml or 50ml. It Automatically recognices the placed syringe.
* The equipment can store up to 2000 historical infusion records, which include the following information: injection mode, injection rate, alarm information, total volume injected, threshold value of the line of occlusion pressure limit, limiting quantity, number of syringe and operation information.
* It has the following alarm functions:
Incorrect position of the syringe.
Residual Limit
Occlusion
End of injection.
No operation
Empty syringe.
Flow limit exceeded.
Plunger decoupled.
Power cord disconnected.
Battery discharged.
Battery exhausted
System error
* Battery operating time: starting from a full charge, the battery can withstand 5 hours of operation at a speed of 5 ml / h on two channels.
* Allows operation and battery charging from the 12V connector of the car.

TECHNICAL SPECIFICATIONS
Dimensions: (Length x Width x Height)
3-channel pump: 469mm x 277mm x 207mm
Weight:
3-channel pump: 5.4kg (includes support).
Feeding:
AC input: 100V ~ to 240V ~, 50/60 Hz.
DC input: 12Vdc
Battery voltage: 12Vdc
Nominal power: 40 VA.
Quality Standards: ISO13485, Disp. A.N.M.A.T. 3266/2013: B.P.F. and Disp. A.N.M.A.T. 2319/02; certified.
Safety Standards: UNE-EN 60601-1. (Eq. IRAM 4220-1 and IEC 601-1), Class I - Prot. Desc. Elect .: CF, UNE-EN 60601-1-2 (Electromagnetic Compatibility).
Authorized by A.N.M.A.T .: PM 1125-18.

ACCESSORIES
● 1 Power cable.
● 1 User manual.
● 1 List of content.
● 1 Certificate of manufacturing quality.
● 1 Warranty card.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Enteral feeding pump
RENT
Includes equipmen rental. For different rental periods, please consult.

Enteral nutrition pump, easy to use, suitable for the infusion of clinical enteral nutrients for example: patients with afagia.
• Convenient:
Transparent pump door, allows visibility of the infusion process.
It operates silently and without vibrations thanks to its high quality motor.
Visual and auditory alarms.
Rinse function of the infusion set.
Con fi guration of infusion time.
• Compact:
Elegant design of the flow line.
Light weight: 1.3 kg.
Easy to transport
• Safe:
Display of the operating status through its indicator.
Programmable fl ow speed and preset volume.
Its design provides an effective prevention against the entry of liquids.
High definition screen, which allows to visualize the main configuration of the infusion process.
Detection of the number of bubbles.
• Smart:
Intermittent feeding output, allowing to establish a period of pause so that the patient can swallow quietly.
It has two modes of operation: Continuous mode and pulse mode.
Stores 1500 historical records.
USB connection
Indication of the status of the power supply.
TECHNICAL SPECIFICATIONS
Flow rate: 1 ml / h ~ 600 ml / h. Preset volume: 0 ml ~ 9999 ml. Total amount accumulated: 1 ml ~ 9999 ml. Accuracy: ± 5%.
Bolus fl ow velocity: 3 adjustable levels: 200 ml / h, 400 ml / h, 600 ml / h.
Automatic Calibration: This unique design can automatically adjust flow accuracy for different viscosity nutrition solutions.
Occlusion pressure range: 500 ± 200 mmHg. Alarm volume: 3 levels.
Keyboard sound: The sound of the key has 3 different levels or it can be muted.
Historical record: Up to 1500 historical records can be stored.
Supply voltage: 110 V ~ A 240 V ~ - 50 Hz or 60 Hz Built-in battery: Ni-Mh, 12 Vdc, 800 mAh.
Operating time: with the battery fully charged, it can run continuously for 5 hours at a speed of 25 ml / h. Power consumption: 11 VA.
Environmental operating conditions:
- Ambient temperature: - 5 ºC to + 40 ºC.
- Ambient humidity: 20% to 90%.
- Environmental pressure: 76.0 kPa to 106.0 kPa. Protection against the entry of liquids: IPX4. Dimensions: 146 mm x 199.5 mm x 114 mm. Weight: 1.3 kg (with hanging support).
Protection against electric shock: Type I, class CF.
ACCESSORIES INCLUDED
- Power cord.
- User's manual.
- Certi fi cate of quality.
- Warranty card.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Defibrillator 3850B
RENT
Includes equipment rental. For different rental periods, please consult.

Feas Electronica Model 3850B Biphasic is an external defibrillator monitor with a rectilinear biphasic waveform of up to 200J, selectable in steps, which allows defibrillation in asynchronous or synchronized mode with the R wave of the patient's ECG for the application of synchronized cardioversion.
It allows the monitoring and surveillance of ECG, with heart rate measurement, through the paddles, or cable to patient with 3 lead or 5 lead.

In addition, optionally, you can monitor pulse oximetry, graphing the plethysmographic curve on the screen and a pulse width indicator bar, indicating the SpO2 and pulse rate values. Both ECG and SpO2 monitoring allow the setting of alarms that allow monitoring of heart rate, SpO2 value and pulse rate.

It is of the portable type, powered by an internal rechargeable battery, a 90 Vac - 240 Vac, 50 Hz or 60 Hz mains supply or an external 12 Vdc source.
Optionally, it has a transcutaneous pacemaker that allows temporary electrical stimulation to the patient in a non-invasive way, in fixed mode (asynchronous) or on demand (synchronous), using electrodes.

Optionally, it can also have a thermal recorder for real-time recording of up to two curves, an ECG lead, and optionally the plethysmographic curve, manually when the operator decides or automatically during charging.

FEATURES

MONITOR
• Electrocardiogram monitor with LCD screen.
• Electrocardiogram signal acquired through 3 or 5 leads patient cable or through paddles.
• Leads selection:
- With 3 lead cable: DI, DII, DIII and PAL.
- With 5 lead cable: DI, DII, DIII, aVR, aVF, aVL, precordial and PAL.
• ECG Gain: 1/6, 1/4, 1/2, 1 and 2 cm / mV.
• ECG waveform cascade.
• Digital indication of heart rate.
• Low battery indicator.
• Luminous Indicator of battery charging.
• Alarms silence key.
• Alarms for Heart Rate from 20 ppm to 250ppm.

DEFIBRILLATOR
• Biphasic waveform.
• Defibrillator use in three steps:
1- Turn on and select energy.
2- Load.
3- Trigger energy discharge by means of push buttons on paddles or from the keyboard.
• Selector switch for charge selection by steps: 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200 Joules.
• Visual and audible indication of charging process completed.
• Stored energy indication on display.
• Energy discharge through paddles by pushing, simultaneously, both pushbuttons on paddles.
• Charge is cancelled internally: When the key DISARM is pressed or when 60 seconds have elapsed since charge process was completed or when selector switch is turned to MONITOR or OFF.
In these cases the energy is discharged internally without reaching the paddles.
• Direct or synchronized discharge.
• Visual and audible indication of synchronized mode.
• Discharge capacity: 150 discharges at maximum energy, with fully charged internal battery.
• Charging time to 200 J, less than 6 seconds.
• The equipment has a power test delivered.

RECORDER (OPTIONAL)
• Direct access key for record command.
• Automatic registration when the power charge process starts up to 20 seconds after pressing the Shock key.
• High resolution thermal array printer.
• Registration speed: 25 mm / s.
• Prints: date and time, energy in joules, Elctrocardiogram waveform, Gain, Heart Rate, Lead, Sweep Speed, Plethysmographics waveform, SPO2, Pulse Rate, Mode, Pacemaker Rate, Pacemaker Current.
• Paper: Width: 50 mm and Diameter Max: 45 mm.
Accessory Included:
50 millimeter roll of paper.

OXYGEN SATURATION (SpO2) (OPTIONAL)
SPO2 measurement:
• Range: 0 to 100%.
• Precision:
Adult ± 2% from 70% and 100%.
Neonatal: ± 3% from 70% and 100%.
In movement: ± 3% from 70% and 100%.
• Resolution: 1%.
• Percentual oxygen saturation alarm:
Upper limit: 1 to100.
Lower limit: 0 to 99.
• Pulse Rate and alarms measurement from 30[1/min] to 250 [1/min] ± 3[1/min] and ± 5 [1/min] with movement.
• Resolution: 1 [1/min].
Accessories included:
Oximetry Sensor Extender
Choice of adult or pediatric, clip-on or disposable oximetry sensor.

TRANSCUTANEOUS PACEMAKER (OPTIONAL)
• External stimulation through electrodes.
• Frequency: 40 ppm to 180 ppm (selectable in steps of 5 ppm).
• Current: 10 mA to 200 mA (selectable in steps of 5 mA).
• Mode: Selectable, on demand or fixed.
• Pulse width: 20 ms ± 5%.
Accessories included:
Elastic belt for external Pacemaker electrode.
Pacemaker lead cable.
Pair of gray electrodes.

TECHNICAL SPECIFICATIONS
• Power input: 90Vac ~ to 240Vac - 50Hz/60Hz or 12 Vcc.
• Rechargeable internal battery, NiMH - 12V, 3500 mAh.
• Power requirement: 180 W (7W during standby).
• Size: 295mm (width) x 200mm (height) x 345mm (length).
• Weight: 6 kg (including external paddles, battery and recorder).

CERTIFICATIONS
• Quality standards:
- CE Marking, Dir. 93/42 / EEC (certificate).
- ISO13485 (certificate).
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.
• Safety rules:
- EN 60601-1 (Eq IRAM 4220-1 and IEC 601-1).
- EN 60601-1-2 (EMC).
- EN 60601-1-6.
- EN 60601-1-8.
- EN 60601-2-4.
- EN 60601-2-27.
- EN 60601-2-49.
- ISO 80601-2-61.
Authorized by the A.N.M.A.T. PM-1125-15.

INCLUDED ACCESSORIES
External paddles for adults and pediatric with coil cable.
Power cable for external +12 power supply.
AC Power cord.
3 or 5 wire cable for patient for ECG.
Disposable electrodes.
User's manual.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - LCD Multiparameter Patient Monitor ECG/RESP/SPO2/PNI
Includes equipment rental. For different rental periods, please consult

Feas Electrónica MULTIPAR LCD Patient Monitor is designed for monitoring the vital signs of patients in Intensive Therapies, Coronary Units, Operating Rooms, Emergency Rooms, and in a wide variety of clinical-hospital environments, where several vital signs need to be monitored of a patient simultaneously.

GENERAL CHARACTERISTICS
FUNCTIONS:
• Menu, messages and instructions in English.
• Single sweep speeds for circulatory and respiratory system: 12.5, 25 and 50 mm/s.
• Visual and audible QRS indicator proceeding from ECG or Oxymeter.
• When QRS proceeds from the Oxymeter, the sound changes its tone as SpO2 modifies.
• Three alarm levels: physiologicals, technicals and generals.
• Upper and lower limits adjustable by the user.
• Visual and audible alarm indicators, on screen and on top of the equipment.
• Hotkey to activate the silent alarms for 2 minutes.
• Alarms Turn Off.
• User friendly menu by using the selector knob and hotkeys.
• Waveform print-out on standard printer.
• Analog outputs for ECG, RESP, IBP and a synchronization pulse.
• Networking Connection for Centralized Monitoring.
• WAIT Mode: stops monitoring, which allows changing to another patient without activating the alarms.
• Screen menu: normal, 7 leads of ECG, Tabular trends or graphical trends, frozen and measuring of calipers, large numbers.

BASIC CHANNELS
ELECTROCARDIOGRAM
• Heart rate: 20 to 300 1/min.
• Gain selection: 1/4, 1/2, 1, 2 and 4 cm/mV, individual or for both channels.
Visual and audible QRS indication.
Automatic selection of 3 or 5 leads patient cable.
• 3 leads with 3 wire patient cable. Lead selection: DI, DII or DIII.
12 leads with 5 wire patient cable. Lead selection: DI, DII, DIII, aVR, aVL, aVF and C1, C2, C3, C4, C5 or C6.
50Hz or 60Hz Notch filter and Diagnostic, Monitoring or Surgery filter.
• Heart rate alarms, minimum and maximum: 20 to 300 1/min and off.
Lead off detection.
Pacemaker pulse detection.
• Pacemaker pulse rejection in counting the heart rate:
•Pulse rejected characteristic:
- Amplitude: from ±2 mV to ±700 mV
- pulse width: from 0.1 ms to 2.0 ms.

ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
• Detects and classifies the following arrhythmias: PVC, PAC, PNC, Bigeminy, Trigeminy, Couple, Run, Tachycardia, Bradycardia, VT, VF and Asystole.
• Store up to 32 events and allows review and printing.
• Measurement of the S-T segment: both the isoelectric point and the ST point are user adjustable (J-point +60 ms or +80 ms, or any other value in ms, from 1 to 120 ms).
• Range: ±2.00 mV.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
• SpO2:
- Range: 35% to 100%
- Precision:
±2% from 80% to 100%.
±3% from 70% to 79%.
- Resolution: 1%.
- Alarm: 0% to 100% and off.
- Initial alarm: 85%.
• Pulse rate:
- Range: 25 to 250 1/min.
- Precision: ±2 1/min.
- Resolution: 1 1/min.
- Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

Accessories included:
Finger pulse oximeter sensor.

RESPIRATORY WAVEFORM
Impedanciometric method.
• Measurement selection: Transabdominal or Transtoraxic.
• Respiratory rate: 5 to 170 1/min.
Apnea Alarm.
• Respiratory rate alarm, minimum and maximum: 5 to 170 1/min and off.

NON-INVASIVE BLOOD PRESSURE (NIBP)
Oscilometric method.
• Use: ADULT or NEONATE.
• Measurement mode: AUTOMATIC or MANUAL.
• Measure unit: mmHg.
• Resolution:
- Pressure: 1 mmHg.
- Pulse rate: 1 1/min.
• Transductor accuracy: ±3 mmHg.
• Ranges:
- PEDIATRIC/ADULT
Systolic pressure: 25 to 265 mmHg.
Diastolic pressure: 12 to 220 mmHg.
Mean pressure: 18 to 245 mmHg.
Pulse rate: 40 to 240 1/min.
- NEONATE
Systolic pressure: 25 to 135 mmHg.
Diastolic pressure: 12 to 110 mmHg.
Mean pressure: 18 to 120 mmHg.
Pulse rate: 40 to 240 1/min.
• Alarm limits of:
- Systolic pressure from 20 to 255 mmHg and off.
- Diastolic pressure from 5 to 220 mmHg and off.
Display of the last 1440 measurements (48 hours, every 2 minutes).
Print-out of the measurements in table form.
Manual or automatic start of the measurement at programmable intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
• Protections:Double protection from over-pressure.
- Over-pressure of 300 mmHg + 20 mmHg for the Adult mode.
- Over-pressure of 150 mmHg + 15 mmHg for the Neonate mode.
- Protection for exceeding time of measurement.

Accessories included:
Cuff for Non Invasive Blood Pressure.
Hose for Non Invasive Blood Pressure.


INVASIVE BLOOD PRESSURE (IBP) (OPTIONAL)
1 or 2 channels of invasive blood pressure, according to customer's choice.
Simultaneous measurement of Systolic, Diastolic and Mean pressure.
• Range: -30 to 300 mmHg.
• Mode selection for Pressure Ranges: Automatic or Manual.
• Mean Pressure alarm: minimum and maximum limits from -30 to 300 mmHg and off.
• Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
• Measure unit: mmHg.
CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

TEMPERATURE (OPTIONAL)
2 channels of temperature.
Compatible with Thermistor Series YSI 400.
Measurement of body temperature or injectate temperature during measurement of Cardiac Output, from -5ºC to 45ºC.
• Resolution: 0.1ºC.
• Accuracy: ±0.2ºC.
• Temperature Alarm, minimum and maximum: -5ºC to 45ºC and off.

Accessories included:
Feas Electrónica temperature sensor - DB9M / G.
Derivation box for Temperature 1 and Temperature 2 parameters, DB9M connector (for two Temperature channels).

CARDIAC OUTPUT (OPTIONAL)
• Entry of patient parameters: Weight, height.
• Selection of:
- Injection volume: 3, 5 and 10 ml.
- Catheter Constant K or Computing Constant (CC)
- Measurement of the temperature in blood and injected.
• Calculation of:
- Body surface.
- CO: Volume in liters / min.
- SV: Stroke volume.
- CI: Cardiac index.
- SI: Systolic index.
Waveform graph for blood temperature as a function of time.
Manual or automatic start of the measurement when injecting.

Accessories included:
Catheter adapter cable for cardiac output - DB9M/G
Derivation box for parameters Temperature 1 and Cardiac Output with DB9.
Derivative box for parameters Temperature 1, Temperature 2 and Cardiac Output with DB9.

SIDESTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 170 1 / min.
• Accuracy:
- CO2, ± 2 mmHg for CO2 <5%.
- <10% of reading for CO2> 5%.
• Resolution:
- CO2, 1 mmHg.
- Respiratory Rate, 1 1 / min.
• Alarm:
- FiCO2, EtCO2, minimum and maximum, 0 to 100 mmHg and off.
- Minimum and maximum Respiratory Rate, 0 to 170 1 / min and off.
Calibration: 2 zero points (ambient air) and gain (5% or 10%).
Compensation for temperature and atmospheric pressure, O2 and N2O.

Accessories included:
Type "T" nozzle for Feás Electrónica lateral flow capnography.
2.4 m patient line for Feás Electrónica lateral flow capnography
Moisture filter for lateral flow capnography, Feas Electrónica

SIDESTREAM CAPNOGRAPHY WITH ANESTHETIC GASES MEASUREMENT (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1/min.
• Accuracy: CO2, ±3%ABS.
• Resolution: CO2, 1 mmHg.
- Respiratory Rate, 1 1/min.
• Alarm: FiCO2, EtCO2 minimum and maximum, 0 to 100 mmHg and off.
• Respiratory Rate minimum and maximum, 0 to 150 1/min and off.
• Measurement of N2 O: from 0% to 100%.
•Automatic detection and measurement of up two anesthetic gases: Halothane, Enflurane,Isoflurane, Sevoflurane and Desflurane.
- Range: 0% to 30%.
- Accuracy: ±8%REL.
- Resolution: 0.1%.
Flowrate:
- 70 to 120 ml/min for neonatal patients.
- 120 to 250 ml/min for adult patients.

Accessories included:
"T" type nozzle for sidestream Capnography, Feas Electrónica.
2,4 m patient line for sidestream Capnography, Feas Electrónica
Water trap for use in lateral flow capnography and anesthetic gases.

MAINSTREAM CAPNOGRAPHY (OPTIONAL)
Measurement of EtCO2, FiCO2 and Respiratory Rate.
• Range:
- CO2, from 0 to 100 mmHg (0% to 13%).
- Respiratory Rate, 4 to 150 1 / min.
• Accuracy: ± 2 mmHg between 0 and 40 mmHg and ± 5% of reading 41 and 70 mmHg.
• Resolution: 1 mmHg.
• Calibration: only zero (ambient air).
Compensation for temperature and atmospheric pressure, O2 and N2O.
Warm-up time <15s, to start measuring, within 2 minutes and obtain measurement values according to specifications.
• Waveform response time: less than 60 ms.

Accessories included:
Mainstream Capnography sensor, Feas Electrónica
Nozzle for mainstream Capnography sensor

FiO2 (OPTIONAL)
Fraction of inspired Oxygen: from 0% to 100%.
Polarographic cell measurement principle.
Alarms minimum and maximum: 0% to 100% and off.

Accessories included:
Derivative box for pressure 1, pressure 2 and / or FiO2 parameters with DB9M.
FiO2 Sensor, Feas Electrónica.
Disposable Oxygen Sensor T-adapter for FiO2, Feas Electrónica
Spiral cable of Oxygen Sensor for FiO2, Feas Electrónica.

ANY OTHER OPTIONAL CHANNEL SPECIFIED CAN BE ADDED IN THE FUTURE.

TECHNICHAL SPECIFICATIONS
15” LCD-TFT Color Monitor
• Resolution: 1024x768.
Wide angle of visibility.
• Dimensions:
- Height: 345mm; Width: 335mm; Depth: 205mm.
• Weight: 6Kg.
• Power Supply: 90V~ to 240V~ (with automatic voltage selection) 50Hz or 60Hz.
• Power Consumption: 85VA.

CERTIFICATIONS
• Quality Standard (Certificates):
- ISO 13485.
- Disp. A.N.M.A.T. 3266/2013: B.P.F.
- Disp. A.N.M.A.T. 2319/02.

• Safety Standard (Certificates):
- UNE-EN60601-1.(Eq. IRAM 4220-1 and IEC 601-1).
- Electrical-shock protection: Class I – CF.
- UNE-EN60601-1-2 (Electromagnetic compatibility).
Authorized by A.N.M.A.T.: PM 1125-2.

ACCESSORIES
Included:3 or 5 leads patient cable.
Disposable electrodes.
SpO2 sensor.
SpO2 extension cable.
Cuff for NIBP.
Silicone hose for NIBP.
Temperature sensor Feas Electrónica - DB9M/G.
Derivation box for parameters of Temperature 1 and Temperature 2, DB9M connector.
FiO2 Sensor Feas Electrónica.
T Adapter for FiO2 sensor Feas Electrónica.
Spiral cable for FiO2 sensor Feas Electrónica.
Power cord.
User manual.
The image is illustrative.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Comen CM300 three channels electrocardiograph
Includes equipment rental. For different rental periods, please consult

CM300 is an Electrocardiograph with an excellent value for money. It is equipped with a thermal printer and an LCD screen. You can view 3 leads simultaneously, print 12 complete leads in groups of 3, and store up to 300 records. Interpretation of results and subsequent printing that will help clinical diagnosis.

Its battery life is approximately 2 hours, so it can be connected to the network or work with the battery itself.
The built-in handle and low weight make it easy to transport. The CM300 is highly recommended because it provides great advanced features at a very low price.

Optional PC software allows you to save all records to PC via USB flash drive, print to common printer, archive as PDF, etc.

CHARACTERISTICS
• Intended to be used in medical institutions such as mobile units.
• Lightweight, robust, compact and transportable, with handle incorporated into the cabinet.
• On-screen display of the electrocardiogram, resolution screen 320 x 240 pixels.
• It contains analysis and diagnostic software.
• Introduction of patient data: name, gender, age, weight, height.
• Digital filtering technology: high precision filter to eliminate interferences of various types and decrease to the deviation of the base line.
• Manual and automatic derivation selection mode.
• Heart rate measurement by selected derivation.
• Printing of three channels or three channels plus heart rate. Various printing formats.
• Thermal printer with 80 mm paper, with selectable print speed and duration.
• Visual alarms for lack of paper, low battery and cable disconnection.
• Storage of 300 ECG records.
• Safe design: ECG amplifier referred to ground, for total patient and operator safety.
• Rechargeable battery with 2 hours of autonomy.
• Optional: connection to data network.

ACCESSORIES INCLUDED WITH THE EQUIPMENT
1 Power cable.
1 patient cable with 10 clips.
6 Chest electrodes.
4 Limb electrodes.
1 Cylinder for paper.
1 Thermosensitive paper roll.
1 Cable to ground.

CERTIFICATES
• Certified Quality Standard:
ISO13485.
Disp. A.N.M.A.T. 3266/2013: B.P.F.
Disp. A.N.M.A.T. 2318/02.
CE certification
• Safety rules:
IEC 60601-1 (Eq IRAM 4220-1 and IEC 601-1- Req.
IEC 60601-1-2 (Electromagnetic Compatibility).
IEC 60601-1-4 (Programmable Medical Devices).
IEC 60601-2-25 (ECG).
IEC 60601-2-51 (ECG registration).

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Odontology implant system
Includes equipment rental. For different rental periods, please consult.

IMPORTANT:The equipment must return in good condition or with wear and tear according to the normal use of the equipment. For the realization of this service, the delivery of a check or promissory note for the value of the equipment will be requested as a guarantee or ask for some other way of guaranteeing.

· It is intended to be used in Dental Implants, in all its stages, and in all types of patients.
· This equipment has been design following international standards, in terms of perfomances and functions.
· High quality motor and great torque, free of vibrations.
· By using a key, you can access the maximum torque selection applied to the tool, between 5 and 45 N cm, or disabled.
· Its adjustable electronic torque meter differentiates it from other equipment.
· Both speed and torque are relative to the selected Reduction. Handpiece reduction: 16:1, 20:1 and 32:1, and digital speed selection, displayed on the display.

MULTIFUNCTION FOOT CONTROL

· Hands-free device for control and programs.
· Heavy multifunction foot control for high stability.
· Main control: It starts the motor in progressive mode (linear) mode or it can be programmed to operate as on/off mode.
· Function: When pressed once, it alternates from one to the other. While holding down, navigate between the three programs of each function.
· Turn: The micro motor tool has the possibility of inverting its rotation, using the DIRECT/REVERSE button on the keyboard, or "Turn" in the multifunction foot control.
· The reverse direction of rotation is indicated on the screen and by an audible tone.
· Pump: Press once, it will turn ON or OFF, when turns ON recalls the last flow rate. Keeping pressed, increase the flow; Off, On 25, On 50, On 75, On 100, Off. This is being represented on the display with 1, 2, 3, or 4 drops.

PERISTALTIC PUMP

Is located in the back of the equipment. Supply 15 ml/min to 45 ml/min for jobs that require it.

CONTENTS

- 1 Main Unit Control.
- 1 Main Power Cable.
- 1 Multifunction Foot Control.
- 1 Brushless Micromotor.
- 1 Sterilisable Headpiece Holder.
- 3 Set Irrigation Tubes.
- 1 User's Manual.
(Handpieces are not included).

TECHNICAL SPECIFICATIONS

Unit Control:
- Global Size: 270mm x 115mm x 230mm.
- Power Input: 90Vac-240Vac 50/60Hz (Automatic).
- Line Fuse: 2.5A/250V, 5mm x 20mm SB.
- Nominal Power: 3W.
- Maximum Power: 150W.
- Screen: LCD - 5.5" (High Luminosity).
- Weight: 2Kg.

Pump:
- Type: Peristaltic with 3 rollers.
- Flow: 15-45ml/min including off.

Multifunction Foot Control:
- Global Size: 18mm x 13.5mm x 43mm.
- Weight: 1.65Kg.
- Protection grade: IPX4.

Motor:
- Rate: 500rpm, 50000 max.
-Torque: 5 to 45 N.cm.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Hemodynamic polygraph
RENT
Includes equipment rental. For different rental periods, please consult.

The Polygraph is a system for conducting electrophysiology and hemodynamic studies.

The system is intended to be used in the acquisition, processing and recording of physiological signals. This includes surface electrocardiography (ECG) and intracavitary (IECG, for Electrophysiology only), heart rate, pulse oximetry (optional), invasive and non-invasive pressure (optional), cardiac output and hemodynamic calculations, now the last two for Hemodynamics only.

It is made up of a central data processing unit (CPU), a preamplifier, a color monitor, a laser printer, a keyboard, a mouse and a recording pedal. For Electrophysiology (if applicable), the Polygraph catheter extension box is added.

FEATURES

Hemodynamics
- 2 simultaneous ECG leads on screen, selectable from 12 leads using 10 electrode wires.
- Up to 4 Invasive Pressures.
- Automatic calculation of the fractional flow reserve (FFR) in real time, without the need for an independent (external) FFR analyzer. The FFR allows determining the hemodynamic severity of vascular lesions.
- Numerical and graphic display of the FFR value.
- Study storage.
- Cardiac output by thermodilution. Calculation of BSA, CO, CI, SV, SI.
- Selection of Catheter Constant K.
- Selection of the injection volume.
- Entry of patient data (height and weight).
- Measurement of the temperature of the blood and the injected.
- Curve of blood temperature variation.
- Messages on screen.
- Recording of studies, stores curves and calculations.
- Measurement of body or injection temperature (during cardiac output).
- Allows you to freeze curves and measure times and amplitudes using calipers.
- Printing of frozen curves or recorded studies.
• HEMODYNAMIC CALCULATIONS:
- Aortic valve area.
- Pulmonary valve area.
- Calculation of pulmonary and systemic vascular resistance.

ELECTROPHYSIOLOGY (OPTIONAL)
- 18 simultaneous endocavitational channels.
- Calipers on Screen: allow you to measure time between events of interest.
- 12 simultaneous surface ECG leads.
- Up to 4 Invasive Pressures.
- Tabular presentation of P.N.I.
- Split Screen: allows you to have a frozen reference image in one window, and in the other view the monitoring in real time.
- Quick access to the curve configuration window, where you can make a continuous selection of position, amplitude, filter selection and customize colors and labels of curves.
- Synchronized triggering of the curve sweep (trigger).
- Recording Studies on a hard disk, to be transported to other computers or personal computers using a pen drive.
- One SpO2 channel.
- Study printing with selection of color and line width.
- Catheters connection box and Cardiac Stimulator
- Review of recorded studies, with the possibility of modifying the gain, position, color and limits (of amplitude) of the curves.
- Quick access keys.

Pulse Oximetry (OPTIONAL)
- Plethysmographic curve, visualization of the oxygen saturation percentage.
- Low perfusion and movement signal measurement.
• SpO2:
- Range from 0% to 100%.
- Adult Accuracy: 2% from 70% to 100%.
- Neonatal Accuracy: 3% from 70% to 100%.
- Accuracy in motion: 3% from 70% to 100%.
- Resolution: 1%.
• Pulse Rate:
- Range from 20 [1 / min] to 300 [1 / min].
- Adult Accuracy: ± 3 [1 / min].
- Neonatal precision: ± 3 [1 / min].
- Accuracy in movement: ± 5 [1 / min].
- Resolution: 1 [1 / min].

Non-Invasive Pressure (OPTIONAL)
- Start of measurement manually or automatically at programmable intervals from 2 minutes to 90 minutes.
- Selection of Adult or Neonatal patient.
- Visualization of the last 1440 measurements.
- Printing of measurements in tabular form.
- Includes cuff.

TECHNICAL SPECIFICATIONS
• Screen: LCD monitor 19 ”
- Dimensions: 19 ”(larger size; optional)
- Resolution: 1024 x 768 dots
•CPU:
- Industrial PC motherboard.
- 160 Gb hard disk.
- Output to connect an additional remote monitor.
• Printing machine:
- Laser Technology.
- Resolution 1200 dpi.
• Power supply:
- 110V ~ / 220V ~.
- Frequency 50/60 Hz.

• Quality Standard:
- ISO9001 (certified).
- ISO13485 (certified).
- Disp. A.N.M.A.T. 3266/2013: B.P.F. (certificate).
- Disp. A.N.M.A.T. 2319/02 (certified).

• Safety Standard:
- UNE-EN60601-1 (Eq. IRAM 4220-1 and IEC 60601-1) Class I.
- Protection against electric shock: CF.
- Authorized by the A.N.M.A.T. PM 1125-23.

ACCESSORIES INCLUDED
- Preamplifier for Hemodynamic Studies.
- Interconnection cable from CPU to Preamplifier.
- Monitor, keyboard and mouse.
- Cable "Y" Minidin-M to 2 Minidin-F.
- Printing machine.
- Recording pedal with cable.
- CPU, monitor and printer power cables.
- Data cable to printer.
- Ground cable for interconnection between CPU and Preamplifier.
- Metal support for Preamplifier.
- Catheter extension box.
- Interconnection cable for catheter extension box.
- Mini equipment holder.
- TMM Bracket Mounting Clamp for TBC.
- Support plate for 4 transducers.
- Power strip with 5 shots.
- Disposable electrodes.
- 4 disposable pressure transducers.
- 4 interconnection cables from Preamplifier to disposable pressure transducers.
- Parameter derivation box for invasive pressure
- Injection temperature sensor.
- Patient cable with 3, 5 or 10 wires.
- Cardiac output catheter adapter cable.
- Oximeter sensor.
- Oximeter extension cable.
- Reus Adult Cuff.
- Silicone hose for PNI, 3m.
- Digital adapter for pressure guides (FFR).
- Volcano Pressure Guide Adapter Cable (FFR).
- Transport trolley (optional).
- User manual.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Electrophysiology Polygraph
RENT
Includes equipment rental. For different rental periods, please consult.

The Polygraph is a system for conducting electrophysiology and hemodynamic studies.

The system is intended to be used in the acquisition, processing and recording of physiological signals. This includes surface electrocardiography (ECG) and intracavitary (IECG, for Electrophysiology only), heart rate, pulse oximetry (optional), invasive and non-invasive pressure (optional), cardiac output and hemodynamic calculations, now the last two for Hemodynamics only.

It is made up of a central data processing unit (CPU), a preamplifier, a color monitor, a laser printer, a keyboard, a mouse and a recording pedal. For Electrophysiology (if applicable), the Polygraph catheter extension box is added.

FEATURES

Hemodynamics
- 2 simultaneous ECG leads on screen, selectable from 12 leads using 10 electrode wires.
- Up to 4 Invasive Pressures.
- Automatic calculation of the fractional flow reserve (FFR) in real time, without the need for an independent (external) FFR analyzer. The FFR allows determining the hemodynamic severity of vascular lesions.
- Numerical and graphic display of the FFR value.
- Study storage.
- Cardiac output by thermodilution. Calculation of BSA, CO, CI, SV, SI.
- Selection of Catheter Constant K.
- Selection of the injection volume.
- Entry of patient data (height and weight).
- Measurement of the temperature of the blood and the injected.
- Curve of blood temperature variation.
- Messages on screen.
- Recording of studies, stores curves and calculations.
- Measurement of body or injection temperature (during cardiac output).
- Allows you to freeze curves and measure times and amplitudes using calipers.
- Printing of frozen curves or recorded studies.
• HEMODYNAMIC CALCULATIONS:
- Aortic valve area.
- Pulmonary valve area.
- Calculation of pulmonary and systemic vascular resistance.

ELECTROPHYSIOLOGY (OPTIONAL)
- 18 simultaneous endocavitational channels.
- Calipers on Screen: allow you to measure time between events of interest.
- 12 simultaneous surface ECG leads.
- Up to 4 Invasive Pressures.
- Tabular presentation of P.N.I.
- Split Screen: allows you to have a frozen reference image in one window, and in the other view the monitoring in real time.
- Quick access to the curve configuration window, where you can make a continuous selection of position, amplitude, filter selection and customize colors and labels of curves.
- Synchronized triggering of the curve sweep (trigger).
- Recording Studies on a hard disk, to be transported to other computers or personal computers using a pen drive.
- One SpO2 channel.
- Study printing with selection of color and line width.
- Catheters connection box and Cardiac Stimulator
- Review of recorded studies, with the possibility of modifying the gain, position, color and limits (of amplitude) of the curves.
- Quick access keys.

Pulse Oximetry (OPTIONAL)
- Plethysmographic curve, visualization of the oxygen saturation percentage.
- Low perfusion and movement signal measurement.
• SpO2:
- Range from 0% to 100%.
- Adult Accuracy: 2% from 70% to 100%.
- Neonatal Accuracy: 3% from 70% to 100%.
- Accuracy in motion: 3% from 70% to 100%.
- Resolution: 1%.
• Pulse Rate:
- Range from 20 [1 / min] to 300 [1 / min].
- Adult Accuracy: ± 3 [1 / min].
- Neonatal precision: ± 3 [1 / min].
- Accuracy in movement: ± 5 [1 / min].
- Resolution: 1 [1 / min].

Non-Invasive Pressure (OPTIONAL)
- Start of measurement manually or automatically at programmable intervals from 2 minutes to 90 minutes.
- Selection of Adult or Neonatal patient.
- Visualization of the last 1440 measurements.
- Printing of measurements in tabular form.
- Includes cuff.

TECHNICAL SPECIFICATIONS
• Screen: LCD monitor 19 ”
- Dimensions: 19 ”(larger size; optional)
- Resolution: 1024 x 768 dots
•CPU:
- Industrial PC motherboard.
- 160 Gb hard disk.
- Output to connect an additional remote monitor.
• Printing machine:
- Laser Technology.
- Resolution 1200 dpi.
• Power supply:
- 110V ~ / 220V ~.
- Frequency 50/60 Hz.

• Quality Standard:
- ISO9001 (certified).
- ISO13485 (certified).
- Disp. A.N.M.A.T. 3266/2013: B.P.F. (certificate).
- Disp. A.N.M.A.T. 2319/02 (certified).

• Safety Standard:
- UNE-EN60601-1 (Eq. IRAM 4220-1 and IEC 60601-1) Class I.
- Protection against electric shock: CF.
- Authorized by the A.N.M.A.T. PM 1125-23.

ACCESSORIES INCLUDED
- Preamplifier for Hemodynamic Studies.
- Interconnection cable from CPU to Preamplifier.
- Monitor, keyboard and mouse.
- Cable "Y" Minidin-M to 2 Minidin-F.
- Printing machine.
- Recording pedal with cable.
- CPU, monitor and printer power cables.
- Data cable to printer.
- Ground cable for interconnection between CPU and Preamplifier.
- Metal support for Preamplifier.
- Catheter extension box.
- Interconnection cable for catheter extension box.
- Mini equipment holder.
- TMM Bracket Mounting Clamp for TBC.
- Support plate for 4 transducers.
- Power strip with 5 shots.
- Disposable electrodes.
- 4 disposable pressure transducers.
- 4 interconnection cables from Preamplifier to disposable pressure transducers.
- Parameter derivation box for invasive pressure
- Injection temperature sensor.
- Patient cable with 3, 5 or 10 wires.
- Cardiac output catheter adapter cable.
- Oximeter sensor.
- Oximeter extension cable.
- Reus Adult Cuff.
- Silicone hose for PNI, 3m.
- Digital adapter for pressure guides (FFR).
- Volcano Pressure Guide Adapter Cable (FFR).
- Transport trolley (optional).
- User manual.

OTHER NAME FOR THE PRODUCT:
Polígrafo/ EPlab Electrophysiology Monitoring and Recording System/ Cathlab Hemodynamic Monitoring and Recording System/ Polígrafo de EP

FREQUENT QUESTION:
Can i add other parameters or optional modules to it in the future?
Yes, you can

IMPORTANT:
The equipment must return in good condition or with wear and tear according to the normal use of the equipment. For the realization of this service, the delivery of a check or promissory note for the value of the equipment will be requested as a guanrantee or ask for some other way of guaranteeing


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Emergency trolley CP 2000
GENERAL CHARACTERISTICS
* It is characterized by its high precision when carrying out low dosages and its stability in the flow rate. It is used particularly for the infusion of sodium nitroprusside, dopamine, propofol and antibiotics for acute and severe diseases.
* The equipment can store customized information of syringes form 33 different manufacturers, Weigao and 32 other manufacturers, configurable by the user.
* The SN-50F66 pump can use 5ml, 10ml, 20ml, 30ml or 50ml sterile and disposable syringes. Automatically recognizes the inserted syringe.
* The set can store up to 2000 infusion history records, including the following information: injection mode, injection speed, alarm information, total volume injected, line occlusion pressure limir threshold value, limiting quantity, number or syringe and information of the operation.
* It has the following alarm functions:
From incorrect position of the syringe.
Of residual limit.
Of occlusion.
End of injection.
No operation.
Of empty syringe.
Flow limit exceeded.
Disengaged plunger.
Power cord disconnected.
Discharged battery.
Depleted battery.
Of system error.
* Battery operating time: from a full charge, the battery can support 5 hours of operation at a rate of 5 ml/h in two channels.
* Allows battery operation and charging from the car´s 12Vdc connector.

TECHNICAL SPECIFICATIONS
Dimensions: (L x W x H)
3 channel pump: 469mm x 277mm x 207mm
Weight:
3-channel pump: 5,4kg (includes support).
Feeding:
AC input: 100V~ to 240V~, 50/60 Hz.
DC input: 12Vcc.
Battery voltage: 12Vcc.
Rated power: 40 VA.
Quality standards: ISO13485, Disp. A.N.M.A.T. 3266/2013: B.P.F. and Disp. A.N.M.A.T. 2319/02; certified.
Safety Standards: UNE-EN 60601-1.(Eq. IRAM 4220-1 e IEC 601-1),Class I - Prot. Desc. Elect.: CF, UNE-EN 60601-1-2 (Electromagnetic Compatibility).
Authorized by the A.N.M.A.T.: PM 1125-18.

ACCESSORIES
● 1 Power cord.
● 1 User´s Manual.
● 1 Content list.
● 1 Manufacturing quality certificate.
● 1 Warranty card.

OTHER PRODUCT NAMES
Volumetric Infusion Pump/ Parenteral Infusion Pump

FREQUENT QUESTIONS
Can it be used on animals?
Yes, it can be used on animals.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Comen CM100 one channel electrocardiograph
Incluye el alquiler del equipo. Para períodos de alquiler diferentes, por favor consulte.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.