Patient Monitor, Multiparameter, of 12,1", Comen, with Batery Li ion include, Touch screen, light, model C80
MULTIPARAMETER MONITOR C80
C80 provides the clinician with advanced functionality and monitoring options to meet the demands of a wide range of care in any hospital setting.
It has a color screen with TFT technology of 10.4" inclined to facilitate observation.
Easy menu use via rotary knob and touch screen.
In addition, it has reserved expansion ports such as for SD card, USB port and network.
Adult, Pediatric, Neonatal Use
ELECTROCARDIOGRAM(ECG)
Heart rate:
•Adult: 15-300 [1/min]
•Neonates/children: 15 - 350 [1/min]
Accuracy: ±1% o ±1 [1/min], whichever is greater.
Resolution: ±1 [1/min]
Gain selection: 1.25mm/mV, 2.5mm/mV, 5.0mm/mV, 10mm/mV, 40mm/mV, Auto.
Manual replacement.
Automatic selection of cable to patient.
3, 5, 12 lead ECG input mode.
Selection of leads:
1. DI, DII, DIII
2. DI, DII, DIII, aVR, aVL, aVF y V
3. DI, DII, DIII, aVR, aVL, aVF y V1 a V6
Frequency response:
•Monitoring mode: 0.5-40Hz.
•Deagnostic mode: 0.05-130Hz.
•Surgery modality: 1-20Hz.
•ST Mode: 0.05-40hZ.
Heart rate alarms:
•Adult: 15 - 300 [1/min]
•Neonates/children: 15 - 350 [1/min]
Loose electrode detection.
Pacemaker Pulse Suppression Capability:
Amplitude of ±2mV~±700mV and pulse width of 0.1ms - 2.0ms.
ANALYSIS AND CLASSIFICATION OF ARRHYTHMIAS
Arrhythmia type: PVC, Asystoloe, VTac, Tachycardia, Ventricular tachycardia, Supraventricular tachycardia, VR, Bigeminia, Trigeminia, R on T, Bradycardia, Ventricular bradycardia, missed beats, Atrial fibrillation, Ventricular fibrillation, Non-captured pacemaker (NCP), Non-captured pacemaker (PNP).
ST segment analysis:
Range: -2mV to +2mV
Accuracy: Measurement tolerance within (-0.8 mV)-(+0.8 mV) is ±0.02 mV or ±10%, whichever is greater. It is not defined for other ranges.
Resolution: 0.01mV (0.1mm)
OXYGEN SATURATION
Plethysmographic curse, arterial oxygen saturation (SpO2) values, peripheral pulse rate (FFPP) and pulse intensity bar.
Comen SpO2:
Range: 0% - 100%
Adult/Child Accuracy: Within 70% -100%, the tolerance is ±2% (without movement).
Neonatal Accuracy: within 70% -100%, the tolerance is ±3% (no movement).
Resolution: 1%
Alarm: 0% to 100%
PULSE RATE
Comen SpO2:
Range: 20 - 254 [1/min]
Resolution: 1[1/min]
Accuracy: ±2 [1/min]
Alarm: 20 a 350 [1/min]
RESPIRATORY CURVE
Impedance method.
Breathing frequency:
•Monitoring range for adult: 0 to 120 [1/min]
•Monitoring range for child/neonate: 0 to 150 [1/min]
Within 7 [1/min] – 150 [1/min], the measurement error is ±2 rpm or ±2%, whichever is greater.
The tolerance is not defined by other ranges.
Alarm Limit Specification:
•Monitoring range for adult: 0 to 120 [1/min]
•Monitoring range for child/neonate: 0 to 150 [1/min]
Apnea alarm
NON-INVASIVE PRESSURE
Automatic Oscillometric Method
Measurement mode: Manual, auto (cyclical) or continuous (not suitable for neonates)
Measurement unit: mmHg
Nominal Monitoring Range:
•Adult Range: Systolic: 40 – 270 mmHg,
Diastolic: 10 – 215 mmHg,
Average: 20 – 235 mmHg
•Pediatric range: Systolic: 40 – 200 mmHg,
Diastolic: 10 – 150 mmHg,
Average: 20 – 165 mmHg
•Neonatal Range: Systolic: 40 – 135 mmHg,
Diastolic: 10 – 100 mmHg,
Average: 20 – 110 mmHg
Alarm Limit Specifications:
•Adult Range: Systolic: 40 – 270 mmHg,
Diastolic: 10 – 215 mmHg,
Average: 20 – 235 mmHg
•Pediatric Range: Systolic: 40 – 200 mmHg,
Diastolic: 10 – 150 mmHg,
Average: 20 – 165 mmHg
•Neonatal Range: Systolic: 40 – 135 mmHg,
Diastolic: 10 – 100 mmHg,
Average: 20 – 110 mmHg
Precision:
The mean error should be less than 5mmHg
The standard deviation should be less than 8mmHg.
Manual or automatic start of measurement at programmable intervals of 1/ 2/ 2.5/ 3/ 4/ 5/ 10/ 15/ 30/ 60/ 90/ 120/ 180/ 240/ 480/ 720min.
Measurement time under continuous mode: 5 min
Overpressure protection:
•Adult: When the cuff pressure exceeds 297 mmHg ±3mmHg, the control valve will relieve the pressure.
•Child: When the cuff pressure exceeds 240 mmHg ±3 mmHg, the control valve will relieve the pressure.
•Neonate: When the cuff pressure exceeds 147 mmHg ±3mmHg, the control valve will relieve the pressure.
Resolution: 1 mmHg.
TEMPERATURE
Measurement Method: direct.
A temperature channel.
Temperature Ranges
Range: 0 - 50° C
Accuracy: ±0.1°C (Exclusive of probe tolerance)
Resolution: 0.1°C
Maximum and minimum temperature alarms:
Range: 0 - 50° C
TECHNICAL SPECIFICATIONS
12.1" color touch screen with TFT technology Resolution: 800 X 600
Inclined screen to facilitate observation.
Dimensions: 344 mm x 291 mm x 165 mm
Weight 3.9 Kg. with battery
Power supply: It has an internal battery and an external AC.
Supply voltage:
100 – 240 Vac, 50/60 Hz.
C80: Standard: Li-Ion 11.8V/2200mAh.
Once fully charged, it ensures continuous autonomous operation for 2 consecutive hours.
Consumption: 60VA
Reserved expansion ports such as SD card, USB port and network.
Quality standards:
IEC 60601-1, IEC 60601-1-8, IEC 60601-2-27, EN 1060-3, IEC 80601-2-30, IEC60601-2-34, IEC60601-2-49, ISO 80601-2-56, ISO 80601-2-61, Class 1, IPX1.