Aggregate of pacemaker module for Desfibrillator 3850B
If you are interested in buying, we suggest you to send us the defibrillator's serial number beforehand to ensure compatibility and verify that your equipment can support the module. When purchasing this module you must send the defibrillator to factory to make the corresponding modifications.

TRANSCUTANEOUS PACEMAKER
- Pacemaker electrode output: Type CF.
- External stimulation through electrodes.
- Frequency: 40 ppm to 180 ppm. (adjustable in steps of 5 ppm).
- Current: 10 mA to 200 mA (adjustable in steps of 5 mA).
- Mode: Selectable, demand or asynchronous.
- Pulse width: 20 ms ± 5%.
- Frequency Accuracy: Within ±5% of displayed value.
-Pacemaker current accuracy: From 10mA to 45mA is ±5mA.
From 50 mA to 200mA is 10%
-Refractory period: 340 ms for frequencies less than or equal to 80 ppm, 240 ms for frequencies greater than 80 ppm.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Aggregate of Recorder module for Desfibrillator 3850B
RECORDER
- Registration hotkey.
- High resolution thermal array printer.
-Resolution
Y axis: 8 points per mm.
X axis: 16 points per mm.
- Paper: width: 50 mm, maximum diameter: 45 mm.
-Printing width: 48 mm.
- Registration speed: 25 mm/s.
- Records the type of patient, the date, the time and in real time, 1 or 2 curves: electrocardiogram and percentage oxygen saturation or cascade electrocardiogram.
-A 50mm x 25m roll of thermal paper is delivered for the UC50 recorder


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Aggregate of blood oxygen saturation module for Desfibrillator 3850B
If you are interested in buying the module, we suggest that you send us the equipment serial number beforehand to ensure compatibility and verify that the module can be added to your equipment. When purchasing this module you must send the equipment to the factory to make the corresponding modifications.

OXYGEN SATURATION
Plethysmographic waveform, display of oxygen saturation (SpO2).
Measurement signal low perfusion and motion.
- SpO2:
Range: 0% to 100%
Adult precision: 2% from 70% to 100%.
Neonate precision: 3% from 70% to 100%.
Motion precision: 3% from 70% to 100%.
Resolution: 1%.
Alarm: 0% to 100% and off.
Initial alarm: 85%.
- Pulse rate:
Range: 20 to 300 1/min.
Adult precision: ±3 1/min.
Neonate precision: ±3 1/min.
Motion precision: ±5 1/min.
Resolution: 1 1/min.
Alarm: 20 to 300 1/min and off.
Waveform gain adjustment x1/2, x1 and x2.

ACCESSORIES
SpO2 sensor, quantity 1.
SpO2 extension cable, quantity 1.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Comen CM100 one channel electrocardiograph
Incluye el alquiler del equipo. Para períodos de alquiler diferentes, por favor consulte.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.