Temporary Cardiac Pacemaker OSYPKA Dual Chamber Model PACE-203
The PACE Model 203 Dual Chamber (DDD) Temporary Cardiac Pacemakeris designed for safe and dependable temporary stimulation of the heart for peri/postoperative pacing and/or therapy for conduction defects and rhythm disturbances.

GENERAL CHARACTERISTICS
• Intuitive dial interface allows for fast and easy access to stimulation and sensitivity settings for the atrium and ventricle.
• High Rate funtion provides rapid atrial stimulation of up to 1,000 ppm (adjustable while applying rapid atrial stimulation).
• Just one button push to automatically measure atrial/ventriuclar rates and intervals, P/R wave amplitudes and AV interval.
• Statistical recording and analysis of pacer activity.
• Standard 9 V alkaline battery provides an average of 240 hrs of continuous operation.
• 2 year warranty.

ACCESSORIES
• Myocardial heartwires for temporary pacing (unipolar, bipolar, and quadripolar)
• Transvenous pacing leads for minimally invasive temporary pacing applications
• Extension cables for temporary pacing leads

TECHNICAL SPECIFICATIONS
Dimensions: 212 mm x 96 mm x 51 mm(8.3” x 3.8” x 2.0”).
Weight 490 g (17.28 oz) including battery
Battery Alkaline: 240 hrs (9.0 V)
Basic Rate 30 - 220 ppm
High Pacing Rate 70 - 1000 ppm
Stimulation Amplitude 0.1 - 18 V
Pulse Width 0.05 - 1.5 ms
Sensitivity Atrium: 0.02 - 20 mV
Ventricular: 1.0 - 20 mV
Refractory Period 250 ms
Runaway Protection 200 ppm ±10 ppm
Pacing Modes:
- Primary: DDD, VVI, AAI, VDD
-Supplementary: D00, V00, A00, DVI
-DAI, VAT, AAT, DDD+AT, DAT




The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
PACE-101 Single Chamber Cardiac Pacemaker OSYPKA
GENERAL CHARACTERISTICS
The PACE 101 single-chamber cardiac pacemaker has the reputation of a legendary workhorse - easy to operate, very durable and light weight. It is intended for temporary atrial or ventricular pacing and supports synchronous (VVI, AAI) and asynchronous (V00, A00) operating modes.

Features:
-Basic stimulation frequency up to 180 ppm.
-Large rotating dials for stimulation frequency, sensitivity, amplitude and pacing mode.
-Rapid atrial pacing up to 720 ppm.
-Long battery lifetime: up to 38 days of operation with a single 9V Alkaline battery.
-Four rotational dials for setup up stimulation frequency, amplitude, sensitivity, and pacing mode.
-Sensed intrinsic activity and paced pulses are indicated optically by a light-emitting diode (LED). Additionally, acoustic signals for sensing and pacing events can be switched on and off.
-Automatic surveillance of battery charge condition. Low battery status is indicated optically by a separate LED labeled 'Low Batt./Error' and acoustically by a warning beep.
-Internal error detection. A device error or defect is indicated by a separate LED labeled 'Low Batt./Error'.
-Run-away protection. The stimulation frequency is limited to a maximum of 180 ppm.
-Protective cover of dials. A protective cover safeguards against accidental changes to the device settings.
Protected terminals in combination with the Series XI Extension Cables ensure touch-proof connection of pacemaker and pacing lead system.

TECHNICAL SPECIFICATIONS
- Dimensions: 6 cm x 11.5 cm x 2.2 cm.
- Weight: 170 g (battery included).
- Alkaline battery: 38 days (72 ppm, 8V).
- Basic stimulation range: 30-180 ppm.
- Rapid stimulation range: basic range x2 or x4 (up to 720 ppm).
- Stimulation amplitude: 0.3 - 12 V.
- Pulse width: 0.75 ms.
- Sensitivity: 1.0 mV at 20 mV, asynchronous.
- Refractory period: 250 ms.
- Protection against uncontrolled running: 200 ± 10 [1 / min].
- Ways of marcapaseo: VVI, VOO, AAI, AOO.

INCLUDED ACCESORIES
- Extension cable.
- User's manual.
- Arm Strap.
- Plastic equipment bag.

OPTIONAL GADGETS
Myocardial heartwires for temporary pacing (unipolar, bipolar)•
Transvenous pacing leads for minimally invasive temporary pacing applications•
Extension cables for temporary pacing leads

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Automatic External Defibrillator (AED) Feas Electrónica, model: Heart+ ResQ NT-381.C
Heart + ResQ NT-381.C is an AED (Automatic External Defibrillator) that provides defibrillation treatments for the rapid resuscitation of the patient in SCA (Sudden Cardiac Arrest). This product is portable, battery operated. It is lightweight and can be operated even by users who have received little or no education / training on the use of the product, through voice prompts that guide the user through the operating sequences.

SCA (Sudden Cardiac Arrest) is the cessation of normal blood circulation due to heart failure. SCA can happen to anyone, anytime, anywhere. Patients with SCA do not show any warning signs or previous symptoms. Some are at higher risk for SCA than others, due to hereditary causes or other reasons.

GENERAL CHARACTERISTICS
Defibrillator
• Waveform: truncated exponential biphasic.
• 150 J (Adult, impedance 50 Ω).
• 50 J (Pediatric, impedance 50 Ω).
Charging time: 9 seconds (± 10 %).
• Shock method: manual button (flashes when ready).
• Mode conversion: Adult/Child toggle button.
(Child Mode delivers 50J Pediatric shock for patients weighing less than 25 Kg, approximately 12 months to 8 years old).

ECG
• ECG record: LEAD II.
• Analysis overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation.
• Analysis time: Approximately 7 seconds.

Battery Check
• Battery reserve level:
High: 3 bars.
Medium: 2 bars.
Low: 1 bar, red status led.
Critical: no bars, red status led.
• Battery status audible warnings.
Critical: Voice prompt states “Replace the battery”.

Data
• Stored data: ECG recordings, operation information (whether power is delivered or not), ambient noise, self diagnostic history.
• Memory capacity:
- ECG: Consecutively make records during analysis and CPR (once in every 5 s, up to 5000 s).
- Audio recording: 60 min in total (4 recording files of 15 min).
- Self-diagnosis: 3000 cases and results.
• Storage type: Internal memory.
• Access/search: “NT-ADV” software (for PC), via Bluethooth.

Audible and Visual Instructions
• After the equipment is switched on, the voice instructions will guide the operator through the resuscitation process, including the positioning of the patches, the indication of “Don’t touch the patient” (during the energy analysis and discharge process) and CPR (Cardio-Pulmonary Resuscitation) instructions.
• The CPR instructions tell the operator when to start Cardio-Pulmonary Resuscitation, according to the AHA Guide.

TECHNICAL SPECIFICATIONS
• Dimensions: 220 mm (W) x 282 mm (L) x 80 mm (H).
• Weight: 2 Kg (including battery and pads).
• Internal non-rechargeable Li-MnO2 battery - 12 Vdc/4200 mAh.
• Protection against ingress of liquids: IPX5 according to IEC60529.
• Temperature:
Storage: 15 ºC to 55 ºC.
Operation: 0 ºC to 45 ºC.
• Humidity:
Storage: 0 % to 95 % (without condensation).
Operation: 0 % to 90 % (without condensation).
• Altitude (Storage/Operation): 0 to 5000 meters.
• Drop test: 1.5 m according to IEC 60068-2-32.
• Quality standards:
ISO 13485 (certified).
Available A.N.M.A.T. 3266/2013: B.P.F.
Available A.N.M.A.T. 2319/02.
• Safety standards:
- IEC 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- IEC 60601-1-2 (Electromagnetic Compatibility).
- IEC 60601-1-6.
- IEC 60601-1-11.
- IEC 60601-2-4.

ACCESSORIES INCLUDED
- Electrode pads
• Shelf life: 2 (two) years.
• Single use.
• Cable length: 2.1 m (± 1 %).
• Applicable ages: All ages (12 months & older).
• Sealing: Airtight.
• Size: 125 mm x 95 mm.
• Pads placement:
Adult: anterior-lateral.
Pediatric: anterior-posterior (see positioning details on the package label).
- 1 Battery Pack
• Type: non-rechargeable Lithium-Manganese Dioxide (Li-MnO2).
• Voltage: 12 Vdc/4200 mAh.
• Standby mode life: Recommended five (5) years at 25 ºC.
• Capacity: 180 shocks @ 25 ºC, 130 shocks at 0 ºC.
• Measures: 22 mm (H) x 164 mm (W) x 54 mm (D).
• Weight: 260 g (±10 %).
- 1 User manual
- 1 bag
- Event Reading Software

OPTIONAL
USB/Bluetooth adapter.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Infrared Thermometer NewTop, model BZ-R6
INFRARED THERMOMETER FOR NON-CONTACT MEASUREMENT.

- Digital forehead thermometer with non-contact infrared technology.
- Easy to use. Measure temperature simply and quickly with the push of a button. The measurement time is less than 1 second.
- It has a sound fever alert (≥ 37.8ºC).
- Ideal for measuring the temperature in adults, children and babies. It can also measure the temperature on surfaces of objects and liquids. The measurement range in people is from 32ºC to 42.9ºC and the range on surfaces is from 0ºC to 100ºC.
- Accurate measurement. Ideal for taking body temperature with a minimum margin of error (± 0.2º).
- LCD screen illuminated in three colors: green, orange and red.
- Safe and hygienic. Ideal for home and professional use.
- Ability to save up to 32 records in its memory.
- Automatic shutdown after 15 seconds.

TECHNICAL CHARACTERISTICS:
Name: Infrared thermometer.
Model: BZ-R6
Non-contact measurement.
Measuring distance: 3-15cm.
Measurement time: ≤0.5s
Audible fever alert (≥ 37.8ºC)
Measurement range on surfaces: from 0 to 100ºC
Measurement range in people: 32-42.5ºC (89.6-108.5ºF)
Precision:
- Within the range 35-42ºC: ± 0.2ºC
- Outside the range 32-42.9º: ± 0.3ºC
Measurement units: ºC / ºF
Memory for the last 32 measurements
Illuminated LCD screen in three colors: green, orange and red.
Automatic shutdown: 15 seconds.
Weight: 142 gr.
Dimensions: 170 x 95 x 45 mm

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Infusion Pump model SN-1500H, Sino MDT Ltd.
Infusion pump of volumetric type with peristaltic perfusion system that accepts universal type set. Comprehensive and effective solution to optimize infusion therapy in all the units of the Hospitals and Health Centers.

• Great quality. Easy to use. Menu in Spanish.
• It has a simple system to calibrate different sets of infusion and can be recorded up to 9 calibrations for different sets.
• It has a horizontal structure with an effective prevention against the entry of liquids. It can be hung from the serum support or placed on a table or car.
• High accuracy, due to an intelligent pulsating compensation made by a double micro-processed unit, resulting in a constant flow of high precision infusion.
• It has five types of infusion mode: Speed ​​mode, Drip mode, Body Weight mode, Time mode and Drug mode.
• Contains drip sensor.
• Contains ultrasonic bubble sensor, high sensitivity.
• It contains a pressure sensor that allows adjustment of 3 levels of occlusion pressure.
• High definition screen, allows the on-screen display of the controlled / running function.
• Illuminated alarm panel (up to 12 alarms). Interface for nurse call.
• Record of 1500 historical events.
• The following features and functions contribute to make the patient's treatment more pleasant:
• Preheating function of the infused solution, with constant temperature of the solution.
• When the pump door is opened, in a dark environment, a light is switched on automatically that allows changing the set or working on the infusion pump without turning on the ambient light.
• Silent operation and without vibrations, because it is equipped with a high quality motor.
• External DC power supply input, also works with 12 VDC input.
• Easy to transport due to its small size and weight (2.5 kg).

TECHNICAL SPECIFICATIONS
Specification of IV set: 20 drops / ml or 60 drops / ml.

Infusion flow rate:
•Set IV of 20 drops / ml:
0.1ml / h ~ 1500ml / h.
1 drops / min ~ 350 drops / min.
• Set IV of 60 drops / ml:
0.1ml / h ~ 200ml / h.
1 drops / min ~ 200 drops / min.
• When the speed is between 0.1ml / h ~ 99.9ml / h, the infusion flow rate is changed from 0.1ml / h.
• When the speed is above 100ml / h, the infusion flow rate is changed from 1ml / h.
• Accuracy: ± 5%.

Bubble Detector:
• Ultrasonic detection method: bubbles are detected above 25μl.

- Occlusion warning speed:
It can be divided into three ranges as indicated,
- High (H) 900mmHg ± 200mmHg (120kPa ± 26,7kPa).
- Central (C) 500mmHg ± 100mmHg (66,7kPa ± 13,3kPa).
- Low (L) 100mmHg ± 50mmHg (13.3kPa ± 6,6kPa).
• Flow rate of the Purge Bolus:
Adjustable between 600ml / h ~ 1000ml / h.
• Kvo rate (to maintain permeable route):
Adjustable minimum flow between 0.1ml / h ~ 5ml / h.

Built-in battery:
• Ni-Mh, 12 Vdc, 2300 mAh, the recharge cycle can be repeated up to 500 times.
• Operating time of the battery: with the battery fully charged, it can run continuously for more than 5 hours at a speed of 25ml / h.

Operation mode:
Keep going.

Protection against the ingress of liquids: IPX4 - Protected against splashing water in any direction.

Historical records:
You can store the historical records of more than 1500 infusions made.

Communication link: SN-1500H: RS-232.

• Supply voltage: 110V ~ to 230V ~ - 50Hz or 60Hz or 12Vcc-5A external.
• Maximum power: 40 VA.
• Classification: class I and internally powered.
• Protection against electric shocks: Type CF.

WORK ENVIRONMENT:
Operating conditions:
• Ambient temperature: + 5ºC to + 40ºC.
• Relative humidity: 20% to 90%.
• Atmospheric pressure: 645mmHg (86kPa) to 795mmHg (106kPa).
Storage conditions:
• Ambient temperature: -5ºC to + 55ºC.
• Relative humidity: 0% to 95%.
• Environmental pressure: 500mmHg (66,7kPa) to 797mmHg (106,3kPa).

ACCESSORIES INCLUDED
1 Power cable to line.
1 user manual.

CERTIFICATES
• Certified Quality Standard:
ISO13485.
Disp. A.N.M.A.T. 3266/2013: B.P.F.
Disp. A.N.M.A.T. 2319/02.
• Safety rules:
IEC 60601-1 (Eq IRAM 4220-1 and IEC 601-1- Req.
IEC 60601-1-2 (Electromagnetic Compatibility).
IEC 60601-2-24 (Infusion Pumps).

FREQUENT QUESTIONS
Do you use any brand of infusion set?
Yes, the pump is not dependent on any brand of supplies. It is recommended that the silicone IV Set be used.

Can it be used on animals?
Yes, it can be used on animals.

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.