Automatic External Defibrillator (AED) Feas Electrónica, model: Heart+ ResQ NT-381.C
Heart + ResQ NT-381.C is an AED (Automatic External Defibrillator) that provides defibrillation treatments for the rapid resuscitation of the patient in SCA (Sudden Cardiac Arrest). This product is portable, battery operated. It is lightweight and can be operated even by users who have received little or no education / training on the use of the product, through voice prompts that guide the user through the operating sequences.

SCA (Sudden Cardiac Arrest) is the cessation of normal blood circulation due to heart failure. SCA can happen to anyone, anytime, anywhere. Patients with SCA do not show any warning signs or previous symptoms. Some are at higher risk for SCA than others, due to hereditary causes or other reasons.

GENERAL CHARACTERISTICS
Defibrillator
• Waveform: truncated exponential biphasic.
• 150 J (Adult, impedance 50 Ω).
• 50 J (Pediatric, impedance 50 Ω).
Charging time: 9 seconds (± 10 %).
• Shock method: manual button (flashes when ready).
• Mode conversion: Adult/Child toggle button.
(Child Mode delivers 50J Pediatric shock for patients weighing less than 25 Kg, approximately 12 months to 8 years old).

ECG
• ECG record: LEAD II.
• Analysis overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation.
• Analysis time: Approximately 7 seconds.

Battery Check
• Battery reserve level:
High: 3 bars.
Medium: 2 bars.
Low: 1 bar, red status led.
Critical: no bars, red status led.
• Battery status audible warnings.
Critical: Voice prompt states “Replace the battery”.

Data
• Stored data: ECG recordings, operation information (whether power is delivered or not), ambient noise, self diagnostic history.
• Memory capacity:
- ECG: Consecutively make records during analysis and CPR (once in every 5 s, up to 5000 s).
- Audio recording: 60 min in total (4 recording files of 15 min).
- Self-diagnosis: 3000 cases and results.
• Storage type: Internal memory.
• Access/search: “NT-ADV” software (for PC), via Bluethooth.

Audible and Visual Instructions
• After the equipment is switched on, the voice instructions will guide the operator through the resuscitation process, including the positioning of the patches, the indication of “Don’t touch the patient” (during the energy analysis and discharge process) and CPR (Cardio-Pulmonary Resuscitation) instructions.
• The CPR instructions tell the operator when to start Cardio-Pulmonary Resuscitation, according to the AHA Guide.

TECHNICAL SPECIFICATIONS
• Dimensions: 220 mm (W) x 282 mm (L) x 80 mm (H).
• Weight: 2 Kg (including battery and pads).
• Internal non-rechargeable Li-MnO2 battery - 12 Vdc/4200 mAh.
• Protection against ingress of liquids: IPX5 according to IEC60529.
• Temperature:
Storage: 15 ºC to 55 ºC.
Operation: 0 ºC to 45 ºC.
• Humidity:
Storage: 0 % to 95 % (without condensation).
Operation: 0 % to 90 % (without condensation).
• Altitude (Storage/Operation): 0 to 5000 meters.
• Drop test: 1.5 m according to IEC 60068-2-32.
• Quality standards:
ISO 13485 (certified).
Available A.N.M.A.T. 3266/2013: B.P.F.
Available A.N.M.A.T. 2319/02.
• Safety standards:
- IEC 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- IEC 60601-1-2 (Electromagnetic Compatibility).
- IEC 60601-1-6.
- IEC 60601-1-11.
- IEC 60601-2-4.

ACCESSORIES INCLUDED
- Electrode pads
• Shelf life: 2 (two) years.
• Single use.
• Cable length: 2.1 m (± 1 %).
• Applicable ages: All ages (12 months & older).
• Sealing: Airtight.
• Size: 125 mm x 95 mm.
• Pads placement:
Adult: anterior-lateral.
Pediatric: anterior-posterior (see positioning details on the package label).
- 1 Battery Pack
• Type: non-rechargeable Lithium-Manganese Dioxide (Li-MnO2).
• Voltage: 12 Vdc/4200 mAh.
• Standby mode life: Recommended five (5) years at 25 ºC.
• Capacity: 180 shocks @ 25 ºC, 130 shocks at 0 ºC.
• Measures: 22 mm (H) x 164 mm (W) x 54 mm (D).
• Weight: 260 g (±10 %).
- 1 User manual
- 1 bag
- Event Reading Software

OPTIONAL
USB/Bluetooth adapter.


The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Training Desfibrillator AED Heart+ResQ NT-381T
A quick guide for the AED training desfibrilator (NT-381 T)

A) Assignment of the identifier
            Press the pediatric mode button and the status button S at the same time.
            When the device turns on, 2 beeps will confirm that it is on and that the process of assigning an identifier has begun.

            1) On the remote control, press the CODE key.
            2) Enter a 2-digit number to assign the desired identifier.
            3) For example 01.02.03 ... 99.
            4) Press ENT to record the entered identifier.

B) Individual control of the DEAs
            Turn on the device and enter the identifier (1 ~ 99) on the remote control and then press the CODE key.

C) Control of several DEAS simultaneously
            Turn on the devices and enter 0 and CODE to simultaneously control these devices regardless of the identifier.

D) Rescue scenarios
            “The team has 11 different scenarios, which are described below.
             Enter the desired scenario number and press the ENT key to save the selection ”.

   Scenario 0 - Defibrillable rhythm
            Press 0 (zero) and then the ENT key to start the scenario.
            The ECG analysis will start automatically after checking the correct placement of the patches 3 times, without pressing any key on the remote control.

   Scenario 1 - Non-defibrillable rhythm
            Press 1 (one) and then the ENT key to start the scenario.
            The ECG analysis will not begin until the instructor presses the NEXT key on the remote control after placing the patches.

   Scenario 2 - Defibrillable rhythm
            Press 2 (two) and then the ENT key to start the scenario.
            The ECG analysis will not begin until the instructor has not pressed the NEXT key on the remote control in the patch placement stage.

   Scenario 3 - Non-defibrillable rhythm
            Press 3 (three) and then the ENT key to start the scenario.
            The ECG analysis will start automatically without inviting to place the patches.

   Scenario 4 - Defibrillable rhythm
            Press 4 (four) and then the ENT key to start the scenario.
            The ECG analysis will start automatically without inviting to place the patches.

   Scenario 5 - Defibrillable rhythm
            Defibrillable rhythm at all times the ECG analysis will start automatically without inviting to place the patches.

   Scenario 6 - CPR (Pulmonary Cardio Resuscitation)

   Scenario 7 - Low battery

   Scenario 8 - An error was detected during the self-diagnosis

   Scenario 9 - Self Diagnosis

   Scenario 10 - PC Mode

The images shown are for illustrative purposes only and may differ from the actual product. Specifications are subject to change without notice.
Monthly rent - Feas Electrónica's Automatic External Defibrillator (AED), model: Heart+ ResQ NT-381.C
GENERAL CHARACTERISTICS
Defibrillator
• Waveform: truncated exponential biphasic.
• 150 J (Adult, impedance 50 Ω).
• 50 J (Pediatric, impedance 50 Ω).
Charging time: 9 seconds (± 10 %).
• Shock method: manual button (flashes when ready).
• Mode conversion: Adult/Child toggle button.
(Child Mode delivers 50J Pediatric shock for patients weighing less than 25 Kg, approximately 12 months to 8 years old).

ECG
• ECG record: LEAD II.
• Analysis overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation.
• Analysis time: Approximately 7 seconds.

Battery Check
• Battery reserve level:
High: 3 bars.
Medium: 2 bars.
Low: 1 bar, red status led.
Critical: no bars, red status led.
• Battery status audible warnings.
Critical: Voice prompt states “Replace the battery”.

Data
• Stored data: ECG recordings, operation information (whether power is delivered or not), ambient noise, self diagnostic history.
• Memory capacity:
- ECG: Consecutively make records during analysis and CPR (once in every 5 s, up to 5000 s).
- Audio recording: 60 min in total (4 recording files of 15 min).
- Self-diagnosis: 3000 cases and results.
• Storage type: Internal memory.
• Access/search: “NT-ADV” software (for PC), via Bluethooth.

Audible and Visual Instructions
• After the equipment is switched on, the voice instructions will guide the operator through the resuscitation process, including the positioning of the patches, the indication of “Don’t touch the patient” (during the energy analysis and discharge process) and CPR (Cardio-Pulmonary Resuscitation) instructions.
• The CPR instructions tell the operator when to start Cardio-Pulmonary Resuscitation, according to the AHA Guide.

TECHNICAL SPECIFICATIONS
• Dimensions: 220 mm (W) x 282 mm (L) x 80 mm (H).
• Weight: 2 Kg (including battery and pads).
• Internal non-rechargeable Li-MnO2 battery - 12 Vdc/4200 mAh.
• Protection against ingress of liquids: IPX5 according to IEC60529.
• Temperature:
Storage: 15 ºC to 55 ºC.
Operation: 0 ºC to 45 ºC.
• Humidity:
Storage: 0 % to 95 % (without condensation).
Operation: 0 % to 90 % (without condensation).
• Altitude (Storage/Operation): 0 to 5000 meters.
• Drop test: 1.5 m according to IEC 60068-2-32.
• Quality standards:
ISO 13485 (certified).
Available A.N.M.A.T. 3266/2013: B.P.F.
Available A.N.M.A.T. 2319/02.
• Safety standards:
- IEC 60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- IEC 60601-1-2 (Electromagnetic Compatibility).
- IEC 60601-1-6.
- IEC 60601-1-11.
- IEC 60601-2-4.

ACCESSORIES INCLUDED
- Electrode pads
• Shelf life: 2 (two) years.
• Single use.
• Cable length: 2.1 m (± 1 %).
• Applicable ages: All ages (12 months & older).
• Sealing: Airtight.
• Size: 125 mm x 95 mm.
• Pads placement:
Adult: anterior-lateral.
Pediatric: anterior-posterior (see positioning details on the package label).
- 1 Battery Pack
• Type: non-rechargeable Lithium-Manganese Dioxide (Li-MnO2).
• Voltage: 12 Vdc/4200 mAh.
• Standby mode life: Recommended five (5) years @ 25 ºC.
• Capacity: 180 shocks @ 25 ºC, 130 shocks @ 0 ºC.
• Measures: 22 mm (H) x 164 mm (W) x 54 mm (D).
• Weight: 260 g (±10 %).
- 1 User manual
- 1 bag
- Event Reading Software

OPTIONAL
USB/Bluetooth adapter.