Defibrillator Monitor 3850B Biphasic for simulation, Feas Electrónica
Simulation defibrillator. DOES NOT DISCHARGE ENERGY.
CHARACTERISTICS MONITOR
- Electrocardiogram monitor with LCD screen.
- Electrocardiogram signal acquired by cable to patient or by paddles.
- Derivations:
- With 3-pin cable: DI, DII, DIII and PAL.
- With 5-pin cable : DI, DII, DIII, aVR, aVF, aVL, Precordial and PAL.
- Cascade electrocardiogram.
- Digital indication of heart rate.
- Low battery indicator.
- Battery charge indicator light.
- Alarm silence key.
- Heart rate alarms from 20 [1/min] to 250 [1/min].
- ECG gain: 1/6, 1/4,1/2, 1 and 2 [cm/mV].
DEFIBRILLATOR
- Use of the defibrillator in three steps:
1. Power on and power selection.
2. Charge.
3. Energy release by means of pushbuttons on paddles or from the keyboard.
- Step load selection key: 2, 3, 5, 7, 10, 20, 50,100, 200, 300 and 360 [Joules].
- Audible and visual indication of end of charge.
- Indication on display of stored energy.
- Energy Trigger by simultaneously pressing both paddle pushbuttons or the “SHOCK” key.
- Internal charge cancellation:
Occurs when the "CANCEL CHARGE" key is pressed or when 60 seconds elapse from the end of the energy charge or when the energy selector key is turned to the "MONITOR" or "OFF" position.
- It does not perform downloads but can, for the simulation, be configured as free or synchronized.
- Audible and visual indication of synchronous mode.
RECORDER
- Direct registration key.
- Automatic registration before an event.
- High resolution thermal recorder.
- Registration speed: 25 mm/s.
- Prints: Date and time, energy in Joules, Electrocardiogram curve, Gain, Heart rate, Lead, Sweep Speed, Plethysmographic curve, SpO2, Pulse Rate, Mode, Pace Rate, Pace Current.
- Paper: Width of 50 mm, Maximun diameter of 45 mm.
TRANSCUTANEOUS PACEMAKER
- External stimulation through electrodes.
- Frequency: 40 ppm to 180 ppm. (Selectable in steps of 5 ppm).
- Current: 10 mA to 200 mA. (Selectable in 5 mA steps).
- Mode: Selectable, on demand of fixed.
- Pulse width: 20 ms ± 5%.
OXYGEN SATURATION (SpO2)
- SpO2 measurement:
- Range 0% a 100%.
- Accuracy:
Adult ±2% between 70% and 100%.
Neonatal ±3% between 70% and 100%.
Moving ±3% between 70% and 100%.
- Resolution: 1%.
- Percentage oxygen saturation alarm:
Upper limit: from 1 to 100.
Lower limit: from 0 to 99.
- Measurement of pulse rate and alarms from 30 to 250 [1/min] ±3 [1/min] and ±5 [1/min] in Movement.
- Resolution: 1 [1/min].
TECHNICAL SPECIFICATIONS
- Supply voltage: 90 Vca a 240 Vca - 50 Hz/60 Hz, or 12 Vcc.
- Internal rechargeable NiMh battery - 12 V, 3500 mAh.
- Consumption: 10 W.
- Size: 295mm (width) x 200mm (height) x 345mm (depth).
- Weight: 5 Kg (Including external paddles, battery and recorder).
ACCESSORIES INCLUDED
- 1 Set of external padles for adults with spiral cable.
- 1 Power cable for external +12 Vdc input.
- 1 Power line cable.
- 1 Patient cable with 3 clips or 5clips for ECG.
- 10 disposable electrodes.
- 1 User Manual.
- 1 Set of electrodes and 1 elastic belt.
- 1 roll of 50 mm paper.
- 1 Oximetry clamp sensor, adult or pediatric.
- 1 Oximetry sensor extension.